Dual source dual energy MDCT: comparison of 80 kVp and weighted average 120 kVp data for conspicuity of hypo-vascular liver metastases
- PMID: 20458250
- DOI: 10.1097/RLI.0b013e3181dfda78
Dual source dual energy MDCT: comparison of 80 kVp and weighted average 120 kVp data for conspicuity of hypo-vascular liver metastases
Abstract
Purpose: To determine whether liver metastases conspicuity is improved at 80 kVp when compared with weighted average (WA) simulated 120 kVp data using dual source dual energy CT.
Methods: A total of 11 patients with 44 hypo-vascular liver metastases underwent contrast enhanced Dual Energy CT (DECT). In all cases the subject's abdominal diameter measured <or=35 cm. Data were reconstructed as a WA of the 140 kVp and 80 kVp acquisitions (simulating 120 kVp) and as a pure 80 kVp data set. A region of interest cursor was placed within the metastasis and adjacent normal parenchyma and attenuation differences and contrast to noise ratios (CNR) were calculated for the metastases at 80 kVp and on the WA acquisition. A mixed model 2-way analysis of variance was used to test whether the attenuation difference between metastases and normal liver was higher at 80 kVp than 120 kVp. An exact Wilcoxon matched-pairs signed rank test was used to test whether CNR was higher at 80 kVp. Cases were retrospectively reviewed to determine whether lesions could be seen on only one or both data sets. As the 80 kVp tube has a smaller detector than the 140 kVp tube, we also noted whether any of the liver lesions were not included on the 80 kVp dataset. Two radiologists in consensus evaluated the 80 kVp data and WA data and subjectively rated hepatic metastases conspicuity on a 4 point scale; with 1 being excellent, 2 good, 3 poor, and 4 not seen.
Results: The mean size of the metastases was 2.6 cm. The mean +/- SD of the attenuation difference between the metastases and the normal liver was 78.37 +/- 24.6 at 80 kVp and 56.89 +/- 17.9 at 120 kVp. The mean difference in attenuation was significantly higher at 80 kVp (P < 0.001). In 2 cases, a metastases was only seen at 80 kVp. The difference between 80 and 120 kVp in terms of CNR was statistically significant (P = 0.042). In one patient, 11 lesions were not included in the smaller field of view of the 80 kVp detector. The conspicuity scores were rated as significantly better at 80 kv than at 120 kVp (P < 0.0001).
Conclusion: When compared with 120 kVp data, pure 80 kVp data acquired from a dual source dual energy MDCT scanner demonstrates greater attenuation differences and improved contrast to noise between metastatic disease and normal liver.
Similar articles
-
Dual-source dual-energy MDCT of pancreatic adenocarcinoma: initial observations with data generated at 80 kVp and at simulated weighted-average 120 kVp.AJR Am J Roentgenol. 2010 Jan;194(1):W27-32. doi: 10.2214/AJR.09.2737. AJR Am J Roentgenol. 2010. PMID: 20028887
-
Dual-energy liver CT: effect of monochromatic imaging on lesion detection, conspicuity, and contrast-to-noise ratio of hypervascular lesions on late arterial phase.AJR Am J Roentgenol. 2014 Sep;203(3):601-6. doi: 10.2214/AJR.13.11337. AJR Am J Roentgenol. 2014. PMID: 25148163
-
Improved Image Quality and Detectability of Hypovascular Liver Metastases on DECT with Different Adjusted Window Settings.Rofo. 2017 Mar;189(3):228-232. doi: 10.1055/s-0042-115571. Epub 2016 Dec 21. Rofo. 2017. PMID: 28002854 English.
-
Dual-energy CT revisited with multidetector CT: review of principles and clinical applications.Diagn Interv Radiol. 2011 Sep;17(3):181-94. doi: 10.4261/1305-3825.DIR.3860-10.0. Epub 2010 Nov 14. Diagn Interv Radiol. 2011. PMID: 20945292 Review.
-
Dual-energy and low-kVp CT in the abdomen.AJR Am J Roentgenol. 2009 Jul;193(1):47-54. doi: 10.2214/AJR.09.2592. AJR Am J Roentgenol. 2009. PMID: 19542394 Free PMC article. Review.
Cited by
-
Value of virtual monochromatic spectral image of dual-layer spectral detector CT with noise reduction algorithm for image quality improvement in obese simulated body phantom.BMC Med Imaging. 2019 Aug 28;19(1):76. doi: 10.1186/s12880-019-0367-8. BMC Med Imaging. 2019. PMID: 31462212 Free PMC article.
-
Assessment of image quality and low-contrast detectability in abdominal CT of obese patients: comparison of a novel integrated circuit with a conventional discrete circuit detector at different tube voltages.Eur Radiol. 2015 Mar;25(3):687-93. doi: 10.1007/s00330-014-3459-4. Epub 2014 Oct 15. Eur Radiol. 2015. PMID: 25316058
-
Application of computed tomography virtual noncontrast spectral imaging in evaluation of hepatic metastases: a preliminary study.Chin Med J (Engl). 2015 Mar 5;128(5):610-4. doi: 10.4103/0366-6999.151656. Chin Med J (Engl). 2015. PMID: 25698191 Free PMC article.
-
Role of dual energy spectral computed tomography in characterization of hepatocellular carcinoma: Initial experience from a tertiary liver care institute.Eur J Radiol Open. 2016 Jul 20;3:162-71. doi: 10.1016/j.ejro.2016.05.007. eCollection 2016. Eur J Radiol Open. 2016. PMID: 27504474 Free PMC article.
-
Material decomposition using iodine quantification on spectral CT for characterising nodules in the cirrhotic liver: a retrospective study.Eur Radiol Exp. 2021 May 28;5(1):22. doi: 10.1186/s41747-021-00220-6. Eur Radiol Exp. 2021. PMID: 34046753 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials