Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years
- PMID: 20458640
- PMCID: PMC2974871
- DOI: 10.1007/s11999-010-1379-4
Highly crosslinked polyethylene reduces wear in total hip arthroplasty at 5 years
Abstract
Background: Although hip arthroplasty reliably relieves pain and improves function, problems have arisen with wear and osteolysis. Highly crosslinked polyethylene has been developed to address this problem although at present there is limited clinical evidence it does so longer term.
Questions/purposes: We compared the in vivo wear of standard versus highly crosslinked polyethylene (HXLP) in primary total hip arthroplasty at a minimum of 5-year followup.
Methods: We enrolled 122 patients in a prospective, double-blinded, randomized trial and followed them annually to assess their progress. Annual radiographs were analyzed using previously validated edge detection software to assess for two-dimensional, three-dimensional, and volumetric wear. The mean follow up was 5.5 years (range, 4.1 to 7 years).
Results: The two-dimensional wear measurements for HXLP showed lower wear compared to the conventional group (0.05 mm/year versus 0.26 mm/year, respectively). Three-dimensional and volumetric wear were similarly lower in the HXLP group.
Conclusions: Highly crosslinked polyethylene undergoes substantially less wear than conventional polyethylene at medium term. The effect of hip arthroplasty longevity will need to be assessed with longer-term studies, but this may lead to a decreased need for revision as a result of less wear and osteolysis.
Figures



Similar articles
-
Oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.Clin Orthop Relat Res. 2015 Dec;473(12):3836-45. doi: 10.1007/s11999-015-4503-7. Clin Orthop Relat Res. 2015. PMID: 26290343 Free PMC article. Clinical Trial.
-
Otto Aufranc Award: Crosslinking Reduces THA Wear, Osteolysis, and Revision Rates at 15-year Followup Compared With Noncrosslinked Polyethylene.Clin Orthop Relat Res. 2018 Feb;476(2):279-290. doi: 10.1007/s11999.0000000000000036. Clin Orthop Relat Res. 2018. PMID: 29529656 Free PMC article. Clinical Trial.
-
Wear and Osteolysis of Highly Crosslinked Polyethylene at 10 to 14 Years: The Effect of Femoral Head Size.Clin Orthop Relat Res. 2016 Feb;474(2):365-71. doi: 10.1007/s11999-015-4319-5. Clin Orthop Relat Res. 2016. PMID: 25903943 Free PMC article.
-
The Otto Aufranc Award. Highly cross-linked polyethylene in total hip arthroplasty: randomized evaluation of penetration rate in cemented and uncemented sockets using radiostereometric analysis.Clin Orthop Relat Res. 2004 Dec;(429):6-16. Clin Orthop Relat Res. 2004. PMID: 15577460 Review.
-
High wear rates and extensive osteolysis in 3 types of uncemented total hip arthroplasty: a review of the PCA, the Harris Galante and the Profile/Tri-Lock Plus arthroplasties with a minimum of 12 years median follow-up in 96 hips.Acta Orthop. 2006 Aug;77(4):575-84. doi: 10.1080/17453670610012638. Acta Orthop. 2006. PMID: 16929433 Review.
Cited by
-
Short-term wear evaluation of thin acetabular liners on 36-mm femoral heads.Clin Orthop Relat Res. 2014 Feb;472(2):624-9. doi: 10.1007/s11999-013-3153-x. Clin Orthop Relat Res. 2014. PMID: 23861047 Free PMC article.
-
Metal-on-conventional polyethylene total hip arthroplasty bearing surfaces have a higher risk of revision than metal-on-highly crosslinked polyethylene: results from a US registry.Clin Orthop Relat Res. 2015 Mar;473(3):1011-21. doi: 10.1007/s11999-014-4105-9. Clin Orthop Relat Res. 2015. PMID: 25560957 Free PMC article.
-
Second-generation annealed highly cross-linked polyethylene exhibits low wear.Clin Orthop Relat Res. 2012 Jun;470(6):1696-704. doi: 10.1007/s11999-011-2177-3. Epub 2011 Dec 9. Clin Orthop Relat Res. 2012. PMID: 22161120 Free PMC article.
-
Poor performance of Enduron polyethylene liner in total hip arthroplasty: a minimum ten-year follow up and ultra-morphological analysis of wear particles.Int Orthop. 2017 Apr;41(4):723-730. doi: 10.1007/s00264-016-3217-0. Epub 2016 May 13. Int Orthop. 2017. PMID: 27178441
-
Dimensional analysis of total hip arthroplasty polyethylenes.Rev Bras Ortop. 2013 Dec 19;48(6):500-504. doi: 10.1016/j.rboe.2013.12.006. eCollection 2013 Nov-Dec. Rev Bras Ortop. 2013. PMID: 31304160 Free PMC article.
References
-
- Bragdon CR, Kwon YM, Geller JA, Greene ME, Freiberg AA, Harris WH, Malchau H. Minimum 6-year followup of highly cross-linked polyethylene in THA. Clin Orthop Relat Res. 2007;465:122–127. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical