Evaluation of the use of real-time PCR for human T cell lymphotropic virus 1 and 2 as a confirmatory test in screening for blood donors
- PMID: 20464136
- DOI: 10.1590/s0037-86822010000200001
Evaluation of the use of real-time PCR for human T cell lymphotropic virus 1 and 2 as a confirmatory test in screening for blood donors
Abstract
Introduction: HTLV-1/2 screening among blood donors commonly utilizes an enzyme-linked immunosorbent assay (EIA), followed by a confirmatory method such as Western blot (WB) if the EIA is positive. However, this algorithm yields a high rate of inconclusive results, and is expensive.
Methods: Two qualitative real-time PCR assays were developed to detect HTLV-1 and 2, and a total of 318 samples were tested (152 blood donors, 108 asymptomatic carriers, 26 HAM/TSP patients and 30 seronegative individuals).
Results: The sensitivity and specificity of PCR in comparison with WB results were 99.4% and 98.5%, respectively. PCR tests were more efficient for identifying the virus type, detecting HTLV-2 infection and defining inconclusive cases.
Conclusions: Because real-time PCR is sensitive and practical and costs much less than WB, this technique can be used as a confirmatory test for HTLV in blood banks, as a replacement for WB.