A multicenter, prospective study of a novel nasal EPAP device in the treatment of obstructive sleep apnea: efficacy and 30-day adherence
- PMID: 20465019
- PMCID: PMC2792968
A multicenter, prospective study of a novel nasal EPAP device in the treatment of obstructive sleep apnea: efficacy and 30-day adherence
Abstract
Study objectives: Evaluate the efficacy of a novel device placed in the nares that imposes an expiratory resistance for the treatment of obstructive sleep apnea (OSA) and evaluate adherence to the device over a 30-day in-home trial period.
Design: One diagnostic and 3 treatment polysomnograms were administered in a Latin-square design to identify the optimal expiratory resistance to be used during the 30-day in-home trial. Subjects had repeat polysomnography with the prescribed device at the end of the 30-day trial.
Setting: Multicenter study.
Participants: Participants (N=34; age 27 to 67) with a baseline apnea-hypopnea index (AHI) > or =5.
Measurements and results: The AHI was reduced from 24.5 _ 23.6 (mean +/- SD) to an average of 13.5 +/- 18.7 (p < 0.001) across initial treatment nights. The AHI was 15.5 +/-+/- 18.9 (p = 0.001) for the prescribed device at the end of the 30-day trial. Of 24 subjects with an AHI > 10 at baseline, 13 achieved an AHI < 10 on the initial treatment nights; 10 had a similar response on the final treatment night. Percent of the night snoring decreased from 27.5 +/- 23.2 to 11.6 +/- 13.7 (p < 0.001) on initial treatment nights and 14.6 +/- 20.6 (p = 0.013) at the end of the trial; Epworth Sleepiness scores decreased from 8.7 +/- 4.0 at baseline to 6.9 +/- 4.4 (p < 0.001) at the end of the trial; the Pittsburgh Sleep Quality Index improved from 7.4 +/- 3.3 to 6.5 +/- 3.6 (p = 0.042). Mean oxygen saturation increased from 94.8 +/- 2.0 to 95.2 +/- 1.9 (p = 0.023) on initial treatment nights and 95.3 +/- 1.9 (p = 0.003) at the end of the trial. Sleep architecture was not affected. Participants reported using the device all night long for 94% of nights during the in-home trial.
Conclusions: Treatment with this novel device was well tolerated and accepted by the participants. An overall reduction in AHI was documented; however, therapeutic response was variable among the participants. Further research is required to identify the ideal candidates for this new therapeutic option in the management of OSA.
Figures
Comment in
-
Auto-PEEP to treat obstructive sleep apnea.J Clin Sleep Med. 2009 Dec 15;5(6):538-9. J Clin Sleep Med. 2009. PMID: 20465020 Free PMC article. No abstract available.
References
-
- McArdle N, Douglas NJ. Effect of continuous positive airway pressure on sleep architecture in the sleep apnea-hypopnea syndrome. Am J Respir Crit Care Med. 2001;164:1459–63. - PubMed
-
- Rosenthal L, Gerhardstein R, Lumley A, Guido P, Day R, Syron ML, Roth T. CPAP therapy in patients with mild OSA: implementation and treatment outcome. Sleep Med. 2000;1:215–20. - PubMed
-
- Campos-Rodriguez F, Perez-Ronchel J, Grilo-Reina A, Lima-Alvarez J, Benitez MA, Almeida-Gonzales C. Long-term effect of continuous positive airway pressure on BP in patients with hypertension and sleep apnea. Chest. 2007;132:1847–52. - PubMed
-
- Kakkar RK, Berry RB. Positive airway pressure treatment for obstructive sleep apnea. Chest. 2007;132:1057–72. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources