. 2010 May;12(5):374-80.
doi: 10.1007/s12094-010-0519-3.

Use of an electron spoiler for radiation treatment of surface skin diseases

Affiliations

Use of an electron spoiler for radiation treatment of surface skin diseases

Víctor Hernández et al. Clin Transl Oncol. 2010 May.

Abstract

Objectives: Low-energy electron beams are characterised by low surface doses with a pronounced dose build-up and penetration of several centimetres, but often a higher surface dose and a lower penetration range is desired. The purpose of this study was to investigate the use of an electron spoiler to modify these beams for treating surface skin diseases and evaluate the feasibility of this method.

Materials and methods: An aluminium foil 4-mm thick covering the end of the electron applicator was used as a spoiler for a 6 MeV electron beam. The dosimetric characteristics of this beam were measured, and Monte Carlo simulations were performed.

Results: The spoiler reduced the practical range and increased surface and build-up doses, but it also significantly broadened the penumbra and increased peripheral doses. Nevertheless, the beam was clinically acceptable when skin collimation with lead was employed. Monte Carlo simulations agreed well with all the experimental measurements.

Conclusions: The feasibility of using a low-energy electron beam with a spoiler for treating surface skin diseases was demonstrated. The method is hygienic and avoids some of the disadvantages associated with the bolus technique, but it is valid only for flat surfaces and perpendicular incidence. As a consequence, it can be an alternative to bolus and other reported methods in certain cases, especially when a particular sterility assurance level is required.

PubMed Disclaimer

References

    1. Phys Med Biol. 2009 Apr 21;54(8):2407-19 - PubMed
    1. Int J Radiat Oncol Biol Phys. 2001 Nov 1;51(3):748-55 - PubMed
    1. Med Phys. 1995 Oct;22(10):1675-83 - PubMed
    1. Radiother Oncol. 1999 Nov;53(2):161-6 - PubMed
    1. N Engl J Med. 2001 Mar 29;344(13):975-83 - PubMed