. 2010 May 17:5:39.
doi: 10.1186/1749-8090-5-39.

Mechanical ventilation and the total artificial heart: optimal ventilator trigger to avoid post-operative autocycling - a case series and literature review

Affiliations

Mechanical ventilation and the total artificial heart: optimal ventilator trigger to avoid post-operative autocycling - a case series and literature review

Allen B Shoham et al. J Cardiothorac Surg. .

Abstract

Many patients with end-stage cardiomyopathy are now being implanted with Total Artificial Hearts (TAHs). We have observed individual cases of post-operative mechanical ventilator autocycling with a flow trigger, and subsequent loss of autocycling after switching to a pressure trigger. These observations prompted us to do a retrospective review of all TAH devices placed at our institution between August 2007 and May 2009. We found that in the immediate post-operative period following TAH placement, autocycling was present in 50% (5/10) of cases. There was immediate cessation of autocycling in all patients after being changed from a flow trigger of 2 L/minute to a pressure trigger of 2 cm H2O. The autocycling group was found to have significantly higher CVP values than the non-autocycling group (P = 0.012). Our data suggest that mechanical ventilator autocycling may be resolved or prevented by the use of a pressure trigger rather than a flow trigger setting in patients with TAHs who require mechanical ventilation.

PubMed Disclaimer

Figures

Figure 1
Figure 1
TAH induced autocycling is present with an actual RR of 28 using a flow trigger.
Figure 2
Figure 2
Autocycling is absent with the use of a pressure trigger as the actual RR = set RR of 16.

Similar articles

Cited by

References

    1. Lloyd-Jones D, Adams R, Carnethon M, De Simone G, Ferguson TB, Flegal K, Ford E, Furie K, Go A, Greenlund K. Heart disease and stroke statistics -- 2009 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation. 2009;119:480–486. doi: 10.1161/CIRCULATIONAHA.108.191259. - DOI - PubMed
    1. Hertz MI, Aurora P, Christie JD, Dobbels F, Edwards LB, Kirk R, Kucheryavaya AY, Rahmel AO, Rowe AW, Taylor DO. Registry of the International Society for Heart and Lung Transplantation: a quarter century of thoracic transplantation. J Heart Lung Transplant. 2008;27:937–942. doi: 10.1016/j.healun.2008.07.019. - DOI - PubMed
    1. Rose EA, Gelijns AC, Moskowitz AJ, Heitjan DF, Stevenson LW, Dembitsky W, Long JW, Ascheim DD, Tierney AR, Levitan RG. Long-term mechanical left ventricular assistance for end-stage heart failure. N Engl J Med. 2001;345:1435–1443. doi: 10.1056/NEJMoa012175. - DOI - PubMed
    1. Gray JNA, Selzman CH. Current status of the total artificial heart. American Heart Journal. 2006;152:4–10. doi: 10.1016/j.ahj.2005.10.024. - DOI - PubMed
    1. Chui PT, Joynt GM, Oh TE, Chui PT, Joynt GM, Oh TE. Severe hyperventilation and respiratory alkalosis during pressure-support ventilation: report of a hazard. Anaesthesia. 1995;50:978–980. doi: 10.1111/j.1365-2044.1995.tb05932.x. - DOI - PubMed

LinkOut - more resources