A new computerized impedance plethysmograph: accuracy in the detection of proximal deep-vein thrombosis in symptomatic outpatients
- PMID: 2048046
A new computerized impedance plethysmograph: accuracy in the detection of proximal deep-vein thrombosis in symptomatic outpatients
Abstract
Because of the lack of specificity of the clinical diagnosis it is appropriate in patients with clinically suspected deep-vein thrombosis to apply an objective test before starting anticoagulant treatment. Impedance plethysmography is a highly accurate technique for the detection of proximal-vein thrombosis with a reported sensitivity and specificity of 93 and 97%, respectively. In all previous reported evaluations of impedance plethysmography an apparatus which was developed in 1971 was used. A new computerized impedance plethysmography, using a novel device to measure impedance, was blindly compared against venography in 443 consecutive outpatients with clinically suspected deep-vein thrombosis. In the first phase of the study the computerized impedance plethysmography test results of 242 symptomatic patients were used to develop a discriminant line. Subsequently, this discriminant line was validated in the second phase of the study in another 201 symptomatic patients. The combined sensitivity and specificity of these two phases for proximal-vein thrombosis was 91% [95% confidence interval (CI), 86 to 94%] and 94% and (95% CI, 90 to 96%), respectively, which compares favourably with impedance plethysmography. It is concluded that computerized impedance plethysmography is a simple, portable, non-invasive technique with a high accuracy for the detection of proximal vein thrombosis. However, before computerized impedance plethysmography can be used as the only test in the diagnosis of deep-vein thrombosis, the safety of withholding anticoagulant treatment to patients with repeated normal computerized test results should be assessed during long-term follow-up studies.
Comment in
-
Noninvasive diagnosis of deep venous thrombosis: lesson from the debate about computerized impedance plethysmography.Thromb Haemost. 1992 Jul 6;68(1):89. Thromb Haemost. 1992. PMID: 1514181 No abstract available.
-
Reliability of non-invasive tests for the diagnosis of venous thrombosis.Thromb Haemost. 1991 Mar 4;65(3):221-2. Thromb Haemost. 1991. PMID: 2048045 No abstract available.
Similar articles
-
Failure of computerized impedance plethysmography in the diagnostic management of patients with clinically suspected deep-vein thrombosis.Thromb Haemost. 1991 Mar 4;65(3):233-6. Thromb Haemost. 1991. PMID: 2048047
-
Detection of deep vein thrombosis with impedance plethysmography and real-time compression ultrasonography in hospitalized patients.Arch Intern Med. 1992 Sep;152(9):1901-3. Arch Intern Med. 1992. PMID: 1520059 Clinical Trial.
-
Reevaluation of the sensitivity of impedance plethysmography for the detection of proximal deep vein thrombosis.Arch Intern Med. 1994 Sep 12;154(17):1930-3. Arch Intern Med. 1994. PMID: 8074596
-
New developments in noninvasive diagnosis of deep vein thrombosis of the lower limbs.Ric Clin Lab. 1990 Jan-Mar;20(1):11-7. doi: 10.1007/BF02910144. Ric Clin Lab. 1990. PMID: 2192431 Review.
-
Factors influencing the reported sensitivity and specificity of impedance plethysmography for proximal deep vein thrombosis.Thromb Haemost. 1994 Nov;72(5):652-8. Thromb Haemost. 1994. PMID: 7900068 Review.
Cited by
-
Meta-analysis of plethysmography and rheography in the diagnosis of deep vein thrombosis.Emerg Med J. 2006 Aug;23(8):630-5. doi: 10.1136/emj.2005.033381. Emerg Med J. 2006. PMID: 16858098 Free PMC article.