Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction
- PMID: 20484423
- DOI: 10.2967/jnumed.109.071571
Quantitative accuracy of clinical 99mTc SPECT/CT using ordered-subset expectation maximization with 3-dimensional resolution recovery, attenuation, and scatter correction
Abstract
We present a calibration method of a clinical SPECT/CT device for quantitative (99m)Tc SPECT. We use a commercially available reconstruction package including ordered-subset expectation maximization (OSEM) with depth-dependent 3-dimensional resolution recovery (OSEM-3D), CT-based attenuation correction, and scatter correction. We validated the method in phantom studies and applied it to images from patients injected with (99m)Tc-diphosponate.
Methods: The following 3 steps were performed to derive absolute quantitative values from SPECT reconstructed images. In step 1, we used simulations to characterize the SPECT/CT system and derive emission recovery values for various imaging parameter settings. We simulated spheres of varying diameters and focused on the dependencies of activity estimation errors on structure size and position, pixel size, count density, and reconstruction parameters. In step 2, we cross-calibrated our clinical SPECT/CT system with the well counter using a large cylinder phantom. This step provided the mapping from image counts to kBq/mL. And in step 3, correction factors from steps 1 and 2 were applied to reconstructed images. We used a cylinder phantom with variable-sized spheres for verification of the method. For in vivo validation, SPECT/CT datasets from 16 patients undergoing (99m)Tc-diphosponate SPECT/CT examinations of the pelvis including the bladder were acquired. The radioactivity concentration in the patients' urine served as the gold standard. Mean quantitative accuracy and SEs were calculated.
Results: In the phantom experiments, the mean accuracy in quantifying radioactivity concentration in absolute terms was within 3.6% (SE, 8.0%), with a 95% confidence interval between -19.4% and +12.2%. In the patient studies, the mean accuracy was within 1.1% (SE, 8.4%), with a 95% confidence interval between -15.4% and +17.5%.
Conclusion: Current commercially available SPECT/CT technology using OSEM-3D reconstruction, scatter correction, and CT-based attenuation correction allows quantification of (99m)Tc radioactivity concentration in absolute terms within 3.6% in phantoms and 1.1% in patients with a focus on the bladder. This opens up the opportunity of SPECT quantitation entering the routine clinical arena. Still, the imprecision caused by unavoidable measurement errors is a dominant factor for absolute quantitation in a clinical setup.
Similar articles
-
Correction of photon attenuation and collimator response for a body-contouring SPECT/CT imaging system.J Nucl Med. 2005 May;46(5):868-77. J Nucl Med. 2005. PMID: 15872362 Clinical Trial.
-
Quantitative CBF measurement using an integrated SPECT/CT system: validation of three-dimensional ordered-subset expectation maximization and CT-based attenuation correction by comparing with O-15 water PET.Ann Nucl Med. 2013 Nov;27(9):822-33. doi: 10.1007/s12149-013-0752-2. Epub 2013 Jul 4. Ann Nucl Med. 2013. PMID: 23824783
-
Hybrid scatter correction applied to quantitative holmium-166 SPECT.Phys Med Biol. 2006 Oct 7;51(19):4773-87. doi: 10.1088/0031-9155/51/19/004. Epub 2006 Sep 8. Phys Med Biol. 2006. PMID: 16985270
-
Review and current status of SPECT scatter correction.Phys Med Biol. 2011 Jul 21;56(14):R85-112. doi: 10.1088/0031-9155/56/14/R01. Epub 2011 Jun 23. Phys Med Biol. 2011. PMID: 21701055 Review.
-
An evidence-based review of quantitative SPECT imaging and potential clinical applications.J Nucl Med. 2013 Jan;54(1):83-9. doi: 10.2967/jnumed.112.111476. J Nucl Med. 2013. PMID: 23283563 Review.
Cited by
-
Accuracy of holmium-166 SPECT/CT quantification over a large range of activities.EJNMMI Phys. 2024 Sep 26;11(1):78. doi: 10.1186/s40658-024-00683-7. EJNMMI Phys. 2024. PMID: 39325204 Free PMC article.
-
Usefulness of Quantitative Bone Single-Photon Emission Computed Tomography/Computed Tomography for Evaluating the Treatment Response of Bone Metastasis in a Lung Cancer Patient.Case Rep Oncol. 2021 Mar 10;14(1):391-396. doi: 10.1159/000513907. eCollection 2021 Jan-Apr. Case Rep Oncol. 2021. PMID: 33776734 Free PMC article.
-
Impact of iterative reconstruction with resolution recovery in myocardial perfusion SPECT: phantom and clinical studies.Sci Rep. 2019 Dec 23;9(1):19618. doi: 10.1038/s41598-019-56097-4. Sci Rep. 2019. PMID: 31873141 Free PMC article.
-
A Novel Method to Suppress the Effect of Subdiaphragmatic Activity in 99mTc Myocardial Perfusion SPECT and Evaluation of Its Usefulness Using a Myocardial Phantom.Ann Nucl Cardiol. 2022;8(1):30-35. doi: 10.17996/anc.22-00154. Epub 2022 Aug 31. Ann Nucl Cardiol. 2022. PMID: 36540175 Free PMC article.
-
The physics of radioembolization.EJNMMI Phys. 2018 Nov 2;5(1):22. doi: 10.1186/s40658-018-0221-z. EJNMMI Phys. 2018. PMID: 30386924 Free PMC article. Review.