Randomized Controlled Trial
. 2010 May;3(5):486-94.
doi: 10.1016/j.jcin.2010.03.007.

Long-term impact of routinely detected early and late incomplete stent apposition: an integrated intravascular ultrasound analysis of the TAXUS IV, V, and VI and TAXUS ATLAS workhorse, long lesion, and direct stent studies

Affiliations
Free article
Randomized Controlled Trial

Long-term impact of routinely detected early and late incomplete stent apposition: an integrated intravascular ultrasound analysis of the TAXUS IV, V, and VI and TAXUS ATLAS workhorse, long lesion, and direct stent studies

Daniel H Steinberg et al. JACC Cardiovasc Interv. 2010 May.
Free article

Abstract

Objectives: We sought to determine the 2-year impact of early and late-acquired incomplete stent apposition (ISA) on clinical events.

Background: The late clinical impact of early or late-acquired ISA in bare-metal stents (BMS) and TAXUS stents (Boston Scientific, Natick, Massachusetts) is debatable.

Methods: We evaluated 1,580 patients enrolled in the intravascular ultrasound (IVUS) substudies of TAXUS IV, V, VI and TAXUS-ATLAS WH, LL, and DS trials.

Results: There were 96 cases of early ISA in 26 (7.2%) BMS patients, 35 (9.7%) TAXUS Express patients (p = 0.28 vs. BMS), and 35 (7.3%) TAXUS Liberté patients (p = 0.21 vs. TAXUS Express, and p = 1.00 vs. BMS). Major adverse cardiovascular events were similar at 9 months in patients with early ISA versus control subjects with no ISA for BMS (3.8% vs. 15.2%, p = 0.13) and for TAXUS (11.6% vs. 8.8%, p = 0.45). There was no impact of early ISA on stent thrombosis. At 9-month follow-up, there were 36 cases of late-acquired ISA in 7 (2.7%) BMS patients, 17 (3.1%) patients with TAXUS slow-release (TAXUS Express or TAXUS Liberté), and 12 (15.4%) patients receiving TAXUS moderate-release. Over 2 ensuing years, major adverse cardiovascular events were similar in patients with late-acquired ISA versus control subjects with no ISA for BMS (14.3% vs. 7.9%, p = 0.54), TAXUS (overall, 8.3% vs. 8.1% p = 0.87), or TAXUS slow-release formulation (0% vs. 7.9%, p = 0.28). There was no impact of late-acquired ISA on stent thrombosis.

Conclusions: Neither routinely detected acute ISA nor routinely detected late-acquired ISA in BMS or TAXUS patients was associated with adverse clinical events over long-term follow-up.