Optical coherence tomography assessment of in vivo vascular response after implantation of overlapping bare-metal and drug-eluting stents
- PMID: 20488410
- DOI: 10.1016/j.jcin.2010.02.008
Optical coherence tomography assessment of in vivo vascular response after implantation of overlapping bare-metal and drug-eluting stents
Abstract
Objectives: We designed a randomized trial exploiting optical coherence tomography (OCT) to assess coverage and apposition of overlapping bare-metal stents (BMS) and drug-eluting stents (DES) in human coronary arteries.
Background: Overlapping DES impair healing in animals. Optical coherence tomography allows accurate in vivo assessment of stent strut coverage and apposition.
Methods: Seventy-seven patients with long coronary stenoses were randomized to overlapping sirolimus-eluting stents (SES), paclitaxel-eluting stents (PES), zotarolimus-eluting stents (ZES), or BMS. The primary goal of the study was to determine the rate of uncovered/malapposed struts in overlap versus nonoverlap segments, according to stent type, at 6-month follow-up with OCT.
Results: A total of 53,047 struts were analyzed. The rate of uncovered/malapposed struts was 1.5 +/- 3.4% and 0.6 +/- 2.7% in overlap versus nonoverlap BMS (p = NS), respectively, and 4.3 +/- 11% and 3.6 +/- 8% in overlap versus nonoverlap DES (p = NS), respectively. There were no differences in the rates of uncovered/malapposed struts between overlapping BMS and DES, likely due to low frequency of uncovered/malapposed struts in ZES (0.1 +/- 0.4%), which offset the higher rates observed in SES (6.7 +/- 9.6%) and PES (6.7 +/- 16.5%, p < 0.05). Overlap segments showed greater neointimal volume obstruction versus nonoverlap segments in all DES (p < 0.05 for all DES types). Strut-level neointimal thickness at overlap and nonoverlap segments were lowest in SES (0.16 +/- 0.1 mm and 0.12 +/- 0.1 mm, respectively) compared with PES (0.27 +/- 0.1 mm and 0.20 +/- 0.1 mm, respectively), ZES (0.40 +/- 0.16 mm and 0.33 +/- 0.13 mm, respectively), and BMS (0.55 +/- 0.31 mm and 0.53 +/- 0.25 mm, respectively, p < 0.05).
Conclusions: As assessed by OCT the impact of DES on vascular healing was similar at overlapping and nonoverlapping sites. However, strut malapposition, coverage pattern, and neointimal hyperplasia differ significantly according to DES type. (Optical Coherence Tomography for Drug Eluting Stent Safety [ODESSA]; NCT00693030).
Copyright 2010 American College of Cardiology Foundation. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Strut coverage and vessel wall response to zotarolimus-eluting and bare-metal stents implanted in patients with ST-segment elevation myocardial infarction: the OCTAMI (Optical Coherence Tomography in Acute Myocardial Infarction) Study.JACC Cardiovasc Interv. 2010 Jun;3(6):680-7. doi: 10.1016/j.jcin.2010.04.005. JACC Cardiovasc Interv. 2010. PMID: 20630463 Clinical Trial.
-
Stent coverage and neointimal proliferation in bare metal stents postdilated with a Paclitaxel-eluting balloon versus everolimus-eluting stents: prospective randomized study using optical coherence tomography at 6-month follow-up.Circ Cardiovasc Interv. 2014 Dec;7(6):760-7. doi: 10.1161/CIRCINTERVENTIONS.113.001146. Epub 2014 Nov 4. Circ Cardiovasc Interv. 2014. PMID: 25371536 Clinical Trial.
-
Evaluation in 3 months duration of neointimal coverage after zotarolimus-eluting stent implantation by optical coherence tomography: the ENDEAVOR OCT trial.JACC Cardiovasc Interv. 2009 Dec;2(12):1240-7. doi: 10.1016/j.jcin.2009.10.006. JACC Cardiovasc Interv. 2009. PMID: 20129551 Clinical Trial.
-
The significance of preclinical evaluation of sirolimus-, paclitaxel-, and zotarolimus-eluting stents.Am J Cardiol. 2007 Oct 22;100(8B):36M-44M. doi: 10.1016/j.amjcard.2007.08.020. Am J Cardiol. 2007. PMID: 17950831 Review.
-
Optical coherence tomography evaluation of intermediate-term healing of different stent types: systemic review and meta-analysis.Eur Heart J Cardiovasc Imaging. 2017 Feb;18(2):159-166. doi: 10.1093/ehjci/jew070. Epub 2016 Apr 20. Eur Heart J Cardiovasc Imaging. 2017. PMID: 27099274
Cited by
-
Adjunctive intra-coronary imaging for the assessment of coronary artery disease.JRSM Cardiovasc Dis. 2016 Jul 8;5:2048004016658142. doi: 10.1177/2048004016658142. eCollection 2016 Jan-Dec. JRSM Cardiovasc Dis. 2016. PMID: 27540480 Free PMC article.
-
Measurement of the blood flow rate and velocity in coronary artery stenosis using intracoronary frequency domain optical coherence tomography: Validation against fractional flow reserve.Int J Cardiol Heart Vasc. 2014 Oct 16;5:68-71. doi: 10.1016/j.ijcha.2014.10.004. eCollection 2014 Dec. Int J Cardiol Heart Vasc. 2014. PMID: 28785616 Free PMC article.
-
Optical Coherence Tomography Imaging: Novel Insights into the Vascular Response After Coronary Stent Implantation.Curr Cardiovasc Imaging Rep. 2012 Aug;5(4):231-238. doi: 10.1007/s12410-012-9138-4. Epub 2012 May 5. Curr Cardiovasc Imaging Rep. 2012. PMID: 22798979 Free PMC article.
-
Automatic stent detection in intravascular OCT images using bagged decision trees.Biomed Opt Express. 2012 Nov 1;3(11):2809-24. doi: 10.1364/BOE.3.002809. Epub 2012 Oct 15. Biomed Opt Express. 2012. PMID: 23162720 Free PMC article.
-
Optical coherence tomography endpoints in stent clinical investigations: strut coverage.Int J Cardiovasc Imaging. 2011 Feb;27(2):271-87. doi: 10.1007/s10554-011-9796-3. Epub 2011 Mar 11. Int J Cardiovasc Imaging. 2011. PMID: 21394615 Free PMC article. Review.
Publication types
MeSH terms
Substances
Associated data
LinkOut - more resources
Full Text Sources
Medical
Research Materials
Miscellaneous