Comparison of macular thickness measurement and segmentation error rate between stratus and fourier-domain optical coherence tomography
Abstract
Background and objective: To compare macular thickness measurements and segmentation error rates between Stratus optical coherence tomography (OCT) (Carl Zeiss Meditec, Inc., Dublin, CA), and Fourier-domain OCT (RTVue, Optovue, Inc., Fremont, CA).
Patients and methods: A retrospective study was performed of 93 normal and pathologic eyes from 79 subjects imaged with both OCT instruments on the same day. Both the macular thickness measurement for each Early Treatment Diabetic Retinopathy Study (ETDRS) zone and the incidence of segmentation error in the central macula between the two instruments were compared.
Results: Macular thickness measurements for all nine ETDRS zones were higher with RTVue compared with Stratus OCT (P < .01). Linear regression analysis showed the highest correlation in the central macula (R(2) = 0.88), with progressively lower correlation peripherally. The overall segmentation error rate was 29% with Stratus OCT versus 32% with RTVue (P > .05).
Conclusion: Macular thickness measurement was greater with RTVue than with Stratus OCT in all ETDRS areas, with the best correlation seen in the central macula. No difference in segmentation error rate was noted between the two instruments.
Copyright 2010, SLACK Incorporated.
Similar articles
-
Comparison of optical coherence tomography models OCT1 and Stratus OCT for macular retinal thickness measurement.Br J Ophthalmol. 2005 Dec;89(12):1581-5. doi: 10.1136/bjo.2005.069815. Br J Ophthalmol. 2005. PMID: 16299134 Free PMC article.
-
Macular thickness measurements in healthy eyes using six different optical coherence tomography instruments.Invest Ophthalmol Vis Sci. 2009 Jul;50(7):3432-7. doi: 10.1167/iovs.08-2970. Epub 2009 Feb 21. Invest Ophthalmol Vis Sci. 2009. PMID: 19234346
-
Artifacts in automatic retinal segmentation using different optical coherence tomography instruments.Retina. 2010 Apr;30(4):607-16. doi: 10.1097/IAE.0b013e3181c2e09d. Retina. 2010. PMID: 20094011
-
Spectral domain optical coherence tomography in diabetic macular edema.Ophthalmic Surg Lasers Imaging. 2009 Nov-Dec;40(6):548-53. doi: 10.3928/15428877-20091030-03. Ophthalmic Surg Lasers Imaging. 2009. PMID: 19928719
-
Artifacts in Macular Optical Coherence Tomography.J Curr Ophthalmol. 2020 Apr 30;32(2):123-131. doi: 10.4103/JOCO.JOCO_83_20. eCollection 2020 Apr-Jun. J Curr Ophthalmol. 2020. PMID: 32671295 Free PMC article. Review.
Cited by
-
Reproducibility of Optovue RTVue Optical Coherence Tomography Retinal Thickness Measurements and Conversion to Equivalent Zeiss Stratus Metrics in Diabetic Macular Edema.Transl Vis Sci Technol. 2015 Jan 22;4(1):5. doi: 10.1167/tvst.4.1.5. eCollection 2015 Jan. Transl Vis Sci Technol. 2015. PMID: 25635237 Free PMC article.
-
Foveal and parafoveal choroidal thickness pattern measuring by swept source optical coherence tomography.Eye (Lond). 2019 Sep;33(9):1443-1451. doi: 10.1038/s41433-019-0404-4. Epub 2019 Apr 8. Eye (Lond). 2019. PMID: 30962543 Free PMC article.
-
Comparison of macular thickness in diabetic macular edema using spectral-domain optical coherence tomography and time-domain optical coherence tomography.J Ophthalmol. 2012;2012:959721. doi: 10.1155/2012/959721. Epub 2012 Jan 18. J Ophthalmol. 2012. PMID: 22315663 Free PMC article.