Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2010 Jun;94(6):720-4.
doi: 10.1136/bjo.2009.164996.

Pattern scan laser photocoagulation: safety and complications, experience after 1301 consecutive cases

Affiliations

Pattern scan laser photocoagulation: safety and complications, experience after 1301 consecutive cases

Raul Velez-Montoya et al. Br J Ophthalmol. 2010 Jun.

Abstract

Background/aims: The pattern scan laser photocoagulator (PASCAL) is a novel laser device that uses 10-30 ms pulse duration for retina photocoagulation. The aim of this study was to analyse the safety profile of this approach.

Methods: This was a retrospective study. We reviewed the clinical records of all laser sessions performed with PASCAL from November 2007 to July 2008. Where there were any complications, we recorded the laser parameters, type, affected retina region, postoperative interval and treatment if required.

Results: There were 1301 consecutive cases. Complications included 17 cases of retinal bleeding (1.3%), two cases of choroidal detachment (0.15%) and one case of exudative retinal detachment (0.07%). There was no statistical difference between the laser parameters used in patients with or without complications.

Conclusion: The laser parameters for PASCAL are safe. The complications and adverse effects encountered in this series are similar to those reported in other studies.

PubMed Disclaimer

Publication types

LinkOut - more resources