Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2010 Jun;63(6):1493-501.
doi: 10.1002/mrm.22347.

Triple repetition time saturation transfer (TRiST) 31P spectroscopy for measuring human creatine kinase reaction kinetics

Affiliations

Triple repetition time saturation transfer (TRiST) 31P spectroscopy for measuring human creatine kinase reaction kinetics

Michael Schär et al. Magn Reson Med. 2010 Jun.

Abstract

Human cardiac phosphorus MR saturation transfer experiments to quantify creatine kinase forward rate constants (k(f)) have previously been performed at 1.5 T. Such experiments could benefit from increased signal-to-noise ratio (SNR) and spectral resolution at 3 T. At 1.5 T, the four-angle saturation transfer method was applied with low-angle adiabatic pulses and surface coils. However, low-angle adiabatic pulses are potentially problematic above 1.5 T due to bandwidth limitations, power requirements, power deposition, and intrapulse spin-spin relaxation. For localized metabolite spin-lattice relaxation time (T(1)) measurements, a dual repetition time approach with adiabatic half-passage pulses was recently introduced to solve these problems at 3 T. Because the saturation transfer experiment requires a T(1) measurement performed while one reacting moiety is saturated, we adapt the dual repetition time approach to measure k(f) using a triple repetition time saturation transfer (TRiST) method. A new pulsed saturation scheme with reduced sensitivity to static magnetic field inhomogeneity and compatibility with cardiac triggering is also presented. TRiST measurements of k(f) are validated in human calf muscle against conventional saturation transfer and found to agree within 3%. The first 3-T TRiST measurements of creatine kinase k(f) in the human calf (n = 6), chest muscle, and heart (n = 8) are 0.26 +/- 0.04 s(-1), 0.23 +/- 0.03 s(-1), and 0.32 +/- 0.07 s(-1), respectively, consistent with prior 1.5 T values.

PubMed Disclaimer

Figures

Figure 1
Figure 1
Scheme of the TRiST acquisitions showing the amplitude modulation of the frequency selective RF saturation versus time. A zoomed version is shown in the inset where 100μs sub-pulses are used for real-time interrogation of the R-wave to enable cardiac triggering. The negative sign of the RF amplitude is to be understood as 180° phase shift. AHP and ACQ denote adiabatic half passage excitation and the acquisition window, respectively.
Figure 2
Figure 2
Amplitude modulation of the DANTE pulse train as shown in Figure 1 broadens the saturated frequency span. (a) The calculated normalized longitudinal magnetization of γ-ATP after 20s of saturation around the saturation frequency with m=5 saturation bands δ=9Hz apart (black) as compared to a single band with m=1 (gray). (b) The saturation bands for m=5 DANTE saturation overlaid over the entire cardiac 31P spectrum. The DANTE sub-pulse separation of τ=0.91ms leads to aliased saturation bands at 1/τ=1100Hz apart.
Figure 3
Figure 3
Results from the Monte Carlo simulations of the TRiST experiment averaged over the kf range 0.1 to 0.4s−1 as a function of TRlong (with γ-ATP saturated) and TRcontrol (control saturation). TRshort was 1.7s (γ-ATP saturated) and NA combinations were chosen for a constant total study time of 38±0.2min. (a) The relative % SD in kf for NA choices that lead to minimum error at each TRlong and TRcontrol combination. (b) The average bias error in kf, corresponding to the TRlong, TRcontrol and NA combinations in part (a). A positive bias of 4% means that the observed kf is less than the true kf by 4%.
Figure 4
Figure 4
Steady-state PCr signal (in arbitrary units, a.u.) measured as a function of TR (diamonds) while γ-ATP is saturated in one volunteer to determine T1,PCr′. T1,PCr′ was determined both by the conventional partial saturation by fitting all 7 data points (dotted line, relaxation curve), and by the 2TR method using only the data points acquired at TRs of 1.5s and 10s (arrows; the grey line is the corresponding implied relaxation curve).
Figure 5
Figure 5
Typical image and TRiST data from a human leg. (a) Axial 1H image annotated with location of the coil center and 1D CSI slice positions. (b–d) 31P spectra extracted from a slice at a depth of 4cm from the coil. The spectra were acquired with: (b) control saturation at a TRcontrol=25.0s to determine M0,PCr; and with saturation of γ-ATP at (c) TRshort=1.5s and (d) TRlong=10.0s to determine M0,PCr′ and T1,PCr′. Note the different vertical scales (arbitrary units, a.u.). Arrows depict the location of the DANTE saturation.
Figure 6
Figure 6
Pseudo-first-order forward rate constants, kf (means±SD) for the CK reaction measured in-vivo in the leg (n=6) and chest (n=8) at 3T. Values plotted from left to right are non-localized (NL) kfs measured by conventional partial saturation (PS) and by TRiST in the calf muscle; kfs measured by 1D CSI localized TRiST as a function of depth in the calf muscle; and kfs measured in the chest and heart by TRiST 1D CSI, as indicated on the horizontal axis.
Figure 7
Figure 7
Typical image and TRiST data from a human heart. (a) Axial 1H image acquired at end-systole and annotated with the location of the coil center and 1D CSI slice positions. (b-d) Cardiac-gated 31P heart spectra extracted from slice 7cm from the coil. The spectra were acquired with: (b) control saturation at a TRcontrol=16.1s to determine M0,PCr; and with saturation of γ-ATP at (c) TRshort=1.7s, and at (d) TRlong=10.0s to determine M0,PCr′ and T1,PCr′. Note the different vertical scales (arbitrary units, a.u.). Arrows depict the location of the DANTE saturation.
Figure 8
Figure 8
Q (means±SD) measured in-vivo in human leg (n-6), chest, and heart (n=8). Values plotted from left to right are from non-localized (NL) studies of calf muscle; 1D CSI studies as a function of depth in the calf; and from 1D CSI studies of the chest and heart, as indicated on the horizontal axis. Q is a measure of spillover of the frequency selective saturation of the γ-ATP resonance onto the PCr resonance.

Similar articles

Cited by

References

    1. Forsen S, Hoffman RA. Study of Moderately Rapid Chemical Exchange Reactions by Means of Nuclear Magnetic Double Resonance. Journal of Chemical Physics. 1963;39(11):2892.
    1. Bottomley PA, Hardy CJ. Mapping Creatine-Kinase Reaction-Rates in Human Brain and Heart with 4-Tesla Saturation Transfer P-31 Nmr. Journal of Magnetic Resonance. 1992;99(2):443–448.
    1. Du F, Zhu XH, Qiao H, Zhang X, Chen W. Efficient in vivo 31P magnetization transfer approach for noninvasively determining multiple kinetic parameters and metabolic fluxes of ATP metabolism in the human brain. Magn Reson Med. 2007;57(1):103–114. - PubMed
    1. Bottomley PA, Ouwerkerk R, Lee RF, Weiss RG. Four-angle saturation transfer (FAST) method for measuring creatine kinase reaction rates in vivo. Magn Reson Med. 2002;47(5):850–863. - PMC - PubMed
    1. Horska A, Fishbein KW, Fleg JL, Spencer RG. The relationship between creatine kinase kinetics and exercise intensity in human forearm is unchanged by age. American journal of physiology. 2000;279(2):E333–339. - PubMed

Publication types

Substances