Continuous glucose monitors and the burden of tight glycemic control in critical care: can they cure the time cost?
- PMID: 20513329
- PMCID: PMC2901040
- DOI: 10.1177/193229681000400317
Continuous glucose monitors and the burden of tight glycemic control in critical care: can they cure the time cost?
Abstract
Background: Tight glycemic control (TGC) in critical care has shown distinct benefits but has also proven to be difficult to obtain. The risk of severe hypoglycemia (<40 mg/dl) raises significant concerns for safety. Added clinical burden has also been an issue. Continuous glucose monitors (CGMs) offer frequent automated measurement and thus the possibility of using them for early detection and intervention of hypoglycemic events. Additionally, regular measurement by CGM may also be able to reduce clinical burden.
Aim: An in silico study investigates the potential of CGM devices to reduce clinical effort in a published TGC protocol.
Methods: This study uses retrospective clinical data from the Specialized Relative Insulin Nutrition Titration (SPRINT) TGC study covering 20 patients from a benchmark cohort. Clinically validated metabolic system models are used to generate a blood glucose (BG) profile for each patient, resulting in 33 continuous, separate BG episodes (6881 patient hours). The in silico analysis is performed with three different stochastic noise models: two Gaussian and one first-order autoregressive. The noisy, virtual CGM BG values are filtered and used to drive the SPRINT TGC protocol. A simple threshold alarm is used to trigger glucose interventions to avert potential hypoglycemia. The Monte Carlo method was used to get robust results from the stochastic noise models.
Results: Using SPRINT with simulated CGM noise, the BG time in an 80-110 mg/dl band was reduced no more than 4.4% to 45.2% compared to glucometer sensors. Antihypoglycemic interventions had negligible effect on time in band but eliminated all recorded hypoglycemic episodes in these simulations. Assuming 4-6 calibration measurements per day, the nonautomated clinical measurements are reduced from an average of 16 per day to as low as 4. At 2.5 min per glucometer measurement, a daily saving of approximately 25-30 min per patient could potentially be achieved.
Conclusions: This paper has analyzed in silico the use of CGM sensors to provide BG input data to the SPRINT TGC protocol. A very simple algorithm was used for early hypoglycemic detection and prevention and tested with four different-sized intravenous glucose boluses. Although a small decrease in time in band (still clinically acceptable) was experienced with the addition of CGM noise, the number of hypoglycemic events was reduced. The reduction to time in band depends on the specific CGM sensor error characteristics and is thus a trade-off for reduced nursing workload. These results justify a pilot clinical trial to verify this study.
(c) 2010 Diabetes Technology Society.
Figures
Similar articles
-
Hypoglycemia detection in critical care using continuous glucose monitors: an in silico proof of concept analysis.J Diabetes Sci Technol. 2010 Jan 1;4(1):15-24. doi: 10.1177/193229681000400103. J Diabetes Sci Technol. 2010. PMID: 20167163 Free PMC article.
-
Stochastic targeted (STAR) glycemic control: design, safety, and performance.J Diabetes Sci Technol. 2012 Jan 1;6(1):102-15. doi: 10.1177/193229681200600113. J Diabetes Sci Technol. 2012. PMID: 22401328 Free PMC article.
-
Pilot study of the SPRINT glycemic control protocol in a Hungarian medical intensive care unit.J Diabetes Sci Technol. 2012 Nov 1;6(6):1464-77. doi: 10.1177/193229681200600628. J Diabetes Sci Technol. 2012. PMID: 23294794 Free PMC article. Clinical Trial.
-
Reduction of number and duration of hypoglycemic events by glucose prediction methods: a proof-of-concept in silico study.Diabetes Technol Ther. 2013 Jan;15(1):66-77. doi: 10.1089/dia.2012.0208. Diabetes Technol Ther. 2013. PMID: 23297671 Review.
-
Using Continuous Glucose Monitoring Data and Detrended Fluctuation Analysis to Determine Patient Condition: A Review.J Diabetes Sci Technol. 2015 Jun 30;9(6):1327-35. doi: 10.1177/1932296815592410. J Diabetes Sci Technol. 2015. PMID: 26134835 Free PMC article. Review.
Cited by
-
Accuracy and stability of an arterial sensor for glucose monitoring in a porcine model using glucose clamp technique.Sci Rep. 2020 Apr 20;10(1):6604. doi: 10.1038/s41598-020-63659-4. Sci Rep. 2020. PMID: 32313062 Free PMC article.
-
Next-generation, personalised, model-based critical care medicine: a state-of-the art review of in silico virtual patient models, methods, and cohorts, and how to validation them.Biomed Eng Online. 2018 Feb 20;17(1):24. doi: 10.1186/s12938-018-0455-y. Biomed Eng Online. 2018. PMID: 29463246 Free PMC article. Review.
-
Glycemic control in critically ill patients.World J Crit Care Med. 2012 Feb 4;1(1):31-9. doi: 10.5492/wjccm.v1.i1.31. eCollection 2012 Feb 4. World J Crit Care Med. 2012. PMID: 24701399 Free PMC article. Review.
-
Untangling glycaemia and mortality in critical care.Crit Care. 2017 Jun 24;21(1):152. doi: 10.1186/s13054-017-1725-y. Crit Care. 2017. PMID: 28645302 Free PMC article.
-
Continuous Glucose Monitoring Measures Can Be Used for Glycemic Control in the ICU: An In-Silico Study.J Diabetes Sci Technol. 2018 Jan;12(1):7-19. doi: 10.1177/1932296817738791. Epub 2017 Nov 6. J Diabetes Sci Technol. 2018. PMID: 29103302 Free PMC article.
References
-
- Capes SE, Hunt D, Malmberg K, Gerstein HC. Stress hyper-glycaemia and increased risk of death after myocardial infarction in patients with and without diabetes: a systematic overview. Lancet. 2000;355(9206):773–778. - PubMed
-
- Finney SJ, Zekveld C, Elia A, Evans TW. Glucose control and mortality in critically ill patients. JAMA. 2003;290(15):2041–2047. - PubMed
-
- Krinsley JS. Association between hyperglycemia and increased hospital mortality in a heterogeneous population of critically ill patients. Mayo Clin Proc. 2003;78(12):1471–1478. - PubMed
-
- McCowen KC, Malhotra A, Bistrian BR. Stress-induced hyper-glycemia. Crit Care Clin. 2001;17(1):107–124. - PubMed
-
- Mizock BA. Alterations in fuel metabolism in critical illness: hyper-glycaemia. Best Pract Res Clin Endocrinol Metab. 2001;15(4):533–551. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical