Parameters influencing increase in pulp chamber temperature with light-curing devices: curing lights and pulpal flow rates
- PMID: 20533637
- DOI: 10.2341/09-234-L
Parameters influencing increase in pulp chamber temperature with light-curing devices: curing lights and pulpal flow rates
Erratum in
- Oper Dent. 2010 Jul-Aug;35(4):487
Abstract
This laboratory study examined the effects of curing lights with different light intensities and changing flow rate on the increase in pulpal temperature during the light curing process and the rate of the subsequent decrease in temperature after the termination of light curing. The tip of a temperature sensor was positioned on the pulpal dentinal wall of the buccal side of the maxillary premolar. Metal tubes were inserted in the palatal and buccal root of the tooth, one for water inflow and the other for water outflow. The tubes were connected to a pump to control the flow rate. The water flow rate was set to 4.2 microl/minute, 28 microl/minute or 70 microl/minute. At each flow rate, the unprepared tooth was light cured from the buccal side 1 mm from the buccalsurface, using four different curing lights. The temperature data were recorded and stored on a computer every second for three minutes. The curing lights that were used were: Astralis 10 (QTH(high), Ivoclar Vivadent), Bluephase 16i (LED(conv), Ivoclar Vivadent) and two experimental LED-curing lights (LED(exp2000), LED(exp3000), Ivoclar Vivadent). The power densities were 1200 mW/cm2, 1600 mW/cm2, 2000 mW/cm2 and 3000 mW/cm2, respectively. The curing lights, LED(conv), LED(exp2000) and LED(exp3000) were activated for 60 seconds, and the QTH(high) was activated for 30 sec- onds. The maximum intrapulpal temperature (TM) and rate of temperature change at 30 seconds after turning off the light (S(30LO)) were analyzed by two-way ANOVA with a post-hoc Tukey test (p < 0.05). The influencing factors were the flow rates and curing lights.
Results: The T(MAX) ranged from 41.0 degrees C to 53.5 degrees C. There was a difference between the curing lights (p < 0.05), with LED(exp3000) > LED(exp2000) > LED(conv) > QTH(high). There was no difference in T(MAX) between the different flow rates (p > 0.05). Both the curing lights and flow rates affected the SE(30LO) (p < 0.05). The S(30LO) was LED(exp3000) < LED(exp2000) > LEDon, , QTH(high) (p < 0.05). The S(30LO) at 70 microl/minutes was higher than at 4.2 pd/minutes and 28 microl/minutes (p < 0.05).
Clinical implication: Because the increase in temperature is directly related to the light intensity and exposure time, curing devices with high power density (> 1200 mW/cm2) should only be activated for a short period of time (< 15 seconds) even in teeth without cavity preparation. The flow rate had only a negligible effect on the temperature increase.
Similar articles
-
Influence of cavity preparation, light-curing units, and composite filling on intrapulpal temperature increase in an in vitro tooth model.Oper Dent. 2014 Sep-Oct;39(5):E195-205. doi: 10.2341/13-068-L. Epub 2014 Apr 10. Oper Dent. 2014. PMID: 24720262
-
Effect of simulated pulpal microcirculation on intrachamber temperature changes following application of various curing units on tooth surface.J Dent. 2009 Jun;37(6):485-90. doi: 10.1016/j.jdent.2009.03.006. Epub 2009 Apr 8. J Dent. 2009. PMID: 19359083
-
Effect of different modes of light curing and resin composites on microleakage of Class II restorations.Odontostomatol Trop. 2008 Dec;31(124):27-34. Odontostomatol Trop. 2008. PMID: 19441264 Clinical Trial.
-
Curing efficiency of three different curing lights at different distances for a hybrid composite.Am J Dent. 2009 Dec;22(6):381-6. Am J Dent. 2009. PMID: 20178217
-
The effectiveness of cure of LED and halogen curing lights at varying cavity depths.Oper Dent. 2003 Nov-Dec;28(6):707-15. Oper Dent. 2003. PMID: 14653284
Cited by
-
Light-curing units used in dentistry: factors associated with heat development-potential risk for patients.Clin Oral Investig. 2017 Jun;21(5):1687-1696. doi: 10.1007/s00784-016-1962-5. Epub 2016 Oct 1. Clin Oral Investig. 2017. PMID: 27695955 Free PMC article.
-
Intraoral Light Curing of Resin Based Material: Adopted by the FDI General Assembly: 27-29 September 2021, Sydney, Australia.Int Dent J. 2022 Feb;72(1):6-7. doi: 10.1016/j.identj.2021.11.005. Int Dent J. 2022. PMID: 35074207 Free PMC article. No abstract available.
-
The effects of extended curing time and radiant energy on microhardness and temperature rise of conventional and bulk-fill resin composites.Clin Oral Investig. 2019 Oct;23(10):3777-3788. doi: 10.1007/s00784-019-02807-1. Epub 2019 Jan 28. Clin Oral Investig. 2019. PMID: 30693403
-
In vitro temperature changes in the pulp chamber caused by laser and Quadwave LED-light curing units.Odontology. 2023 Jul;111(3):668-679. doi: 10.1007/s10266-022-00780-y. Epub 2022 Dec 19. Odontology. 2023. PMID: 36534223
-
High-Power LED Units Currently Available for Dental Resin-Based Materials-A Review.Polymers (Basel). 2021 Jun 30;13(13):2165. doi: 10.3390/polym13132165. Polymers (Basel). 2021. PMID: 34208978 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources