. 2011 Jan;469(1):180-7.
doi: 10.1007/s11999-010-1424-3. Epub 2010 Jun 11.

Do ion levels in hip resurfacing differ from metal-on-metal THA at midterm?

Affiliations
Comparative Study

Do ion levels in hip resurfacing differ from metal-on-metal THA at midterm?

A Moroni et al. Clin Orthop Relat Res. 2011 Jan.

Abstract

Background: Metal-on-metal Birmingham hip resurfacing (MOM-BHR) is an alternative to metal-on-metal total hip arthroplasty (MOM-THA), especially for young and/or active patients. However, wear resulting in increased serum ion levels is a concern.

Questions/purposes: We asked whether (1) serum chromium (Cr), cobalt (Co), and molybdenum (Mo) concentrations would differ between patients with either MOM-BHR or MOM-THA at 5 years, (2) confounding factors such as gender would influence ion levels; and (3) ion levels would differ at 2 and 5 years for each implant type.

Patients and methods: Ions were measured in two groups with either MOM-BHR (n = 20) or MOM-THA (n = 35) and a mean 5-year followup, and two groups with either MOM-BHR (n = 15) or MOM-THA (n = 25) and a mean 2-year followup. Forty-eight healthy blood donors were recruited for reference values.

Results: At 5 years, there were no differences in ion levels between patients with MOM-BHR or MOM-THA. Gender was a confounding factor, and in the MOM-BHR group at 5 years, Cr concentrations were greater in females compared with those of males. Mean ion levels were similar in patients with 2 and 5 years of followup for each implant type. Ion levels in patients were sevenfold to 10-fold higher than in controls.

Conclusions: As the metal ion concentrations in the serum at 5 years were in the range reported in the literature, we do not believe concerns regarding excessive metal ion levels after MOM-BHR are justified.

Level of evidence: Level III, therapeutic study. See the Guidelines for Authors for a complete description of level of evidence.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Amstutz HC, Le Duff MJ. Background of metal-on-metal resurfacing. Proc Inst Mech Eng H. 2006;220:85–94. - PubMed
    1. Anissian HL, Stark A, Good V, Dahlstrand H, Clarke IC. The wear pattern in metal-on-metal hip prostheses. J Biomed Mater Res. 2001;58:673–678. doi: 10.1002/jbm.1068. - DOI - PubMed
    1. Back DL, Young DA, Shimmin AJ. How do serum cobalt and chromium levels change after metal-on-metal hip resurfacing? Clin Orthop Relat Res. 2005;438:177–181. doi: 10.1097/01.blo.0000166901.84323.5d. - DOI - PubMed
    1. Black J. In vivo corrosion of a cobalt-base and its biological consequences. In: Hildebrand HF, Champy M eds. Biocompatibility of Co-Cr-Ni Alloys. NATO ASI Series. New York, NY: Plenum Press; 1988;83–100.
    1. Clarke MT, Lee PT, Arora A, Villar RN. Levels of metal ions after small- and large-diameter metal-on-metal hip arthroplasty. J Bone Joint Surg Br. 2003;85:913–917. - PubMed

Publication types

MeSH terms