Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction

Affiliations

Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction

Maria Raissaki et al. Pediatr Radiol. 2010 Nov.

Abstract

Background: CT scans of the brain, sinuses and petrous bones performed as the initial imaging test for a variety of indications have the potential to expose the eye-lens, considered among the most radiosensitive human tissues, to a radiation dose. There are several studies in adults discussing the reduction of orbital dose resulting from the use of commercially available bismuth-impregnated latex shields during CT examinations of the head.

Objective: To evaluate bismuth shielding-induced artefacts and to provide suggestions for optimal eye-lens shielding in paediatric head CT.

Materials and methods: A bismuth shield was placed over the eyelids of 60 consecutive children undergoing head CT. Images were assessed for the presence and severity of artefacts with regard to eye-shield distance and shield wrinkling. An anthropomorphic paediatric phantom and thermoluminescence dosimeters (TLDs) were used to study the effect of eye lens-to-shield distance on shielding efficiency.

Results: Shields were tolerated by 56/60 children. Artefacts were absent in 45% of scans. Artefacts on orbits, not affecting and affecting orbit evaluation were noted in 39% and 14% of scans, respectively. Diagnostically insignificant artefacts on intracranial structures were noted in 1 case (2%) with shield misplacement. Mean eye-lens-to-shield distance was 8.8 mm in scans without artefacts, and 4.3 mm and 2.2 mm in scans with unimportant and diagnostically important artefacts, respectively. Artefacts occurred in 8 out of 9 cases with shield wrinkling. Dose reduction remained unchanged for different shield-to-eye distances.

Conclusion: Bismuth shielding-related artefacts occurring in paediatric head CT are frequent, superficial and diagnostically insignificant when brain pathology is assessed. Shields should be placed 1 cm above the eyes when orbital pathology is addressed. Shield wrinkling should be avoided.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Med Phys. 2008 Oct;35(10):4567-76 - PubMed
    1. AJNR Am J Neuroradiol. 2001 Jun-Jul;22(6):1194-8 - PubMed
    1. Radiology. 2008 May;247(2):499-506 - PubMed
    1. Med Phys. 2005 Jun;32(6):1621-9 - PubMed
    1. Eur Radiol. 2006 Oct;16(10):2334-40 - PubMed

LinkOut - more resources