Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction
- PMID: 20552187
- DOI: 10.1007/s00247-010-1715-6
Eye-lens bismuth shielding in paediatric head CT: artefact evaluation and reduction
Abstract
Background: CT scans of the brain, sinuses and petrous bones performed as the initial imaging test for a variety of indications have the potential to expose the eye-lens, considered among the most radiosensitive human tissues, to a radiation dose. There are several studies in adults discussing the reduction of orbital dose resulting from the use of commercially available bismuth-impregnated latex shields during CT examinations of the head.
Objective: To evaluate bismuth shielding-induced artefacts and to provide suggestions for optimal eye-lens shielding in paediatric head CT.
Materials and methods: A bismuth shield was placed over the eyelids of 60 consecutive children undergoing head CT. Images were assessed for the presence and severity of artefacts with regard to eye-shield distance and shield wrinkling. An anthropomorphic paediatric phantom and thermoluminescence dosimeters (TLDs) were used to study the effect of eye lens-to-shield distance on shielding efficiency.
Results: Shields were tolerated by 56/60 children. Artefacts were absent in 45% of scans. Artefacts on orbits, not affecting and affecting orbit evaluation were noted in 39% and 14% of scans, respectively. Diagnostically insignificant artefacts on intracranial structures were noted in 1 case (2%) with shield misplacement. Mean eye-lens-to-shield distance was 8.8 mm in scans without artefacts, and 4.3 mm and 2.2 mm in scans with unimportant and diagnostically important artefacts, respectively. Artefacts occurred in 8 out of 9 cases with shield wrinkling. Dose reduction remained unchanged for different shield-to-eye distances.
Conclusion: Bismuth shielding-related artefacts occurring in paediatric head CT are frequent, superficial and diagnostically insignificant when brain pathology is assessed. Shields should be placed 1 cm above the eyes when orbital pathology is addressed. Shield wrinkling should be avoided.
Similar articles
-
Use of bismuth shield for protection of superficial radiosensitive organs in patients undergoing computed tomography: a literature review and meta-analysis.Radiol Phys Technol. 2019 Mar;12(1):6-25. doi: 10.1007/s12194-019-00500-2. Epub 2019 Feb 21. Radiol Phys Technol. 2019. PMID: 30790174 Review.
-
Reduction of eye lens radiation dose by orbital bismuth shielding in pediatric patients undergoing CT of the head: a Monte Carlo study.Med Phys. 2005 Apr;32(4):1024-30. doi: 10.1118/1.1881852. Med Phys. 2005. PMID: 15895586
-
[Protection of eye lens in computed tomography--dose evaluation on an anthropomorphic phantom using thermo-luminescent dosimeters and Monte-Carlo simulations].Rofo. 2008 Dec;180(12):1047-53. doi: 10.1055/s-2008-1027814. Epub 2008 Nov 28. Rofo. 2008. PMID: 19235699 German.
-
Reducing absorbed dose to eye lenses in head CT examinations: the effect of bismuth shielding.Australas Phys Eng Sci Med. 2016 Jun;39(2):583-9. doi: 10.1007/s13246-016-0445-y. Epub 2016 Apr 20. Australas Phys Eng Sci Med. 2016. PMID: 27098155
-
Radiation dose and risk to the lens of the eye during CT examinations of the brain.J Med Imaging Radiat Oncol. 2019 Dec;63(6):786-794. doi: 10.1111/1754-9485.12950. Epub 2019 Sep 13. J Med Imaging Radiat Oncol. 2019. PMID: 31520467 Review.
Cited by
-
Use of bismuth shield for protection of superficial radiosensitive organs in patients undergoing computed tomography: a literature review and meta-analysis.Radiol Phys Technol. 2019 Mar;12(1):6-25. doi: 10.1007/s12194-019-00500-2. Epub 2019 Feb 21. Radiol Phys Technol. 2019. PMID: 30790174 Review.
-
Radiation dose reduction in parasinus CT by spectral shaping.Neuroradiology. 2017 Feb;59(2):169-176. doi: 10.1007/s00234-016-1780-0. Epub 2017 Jan 16. Neuroradiology. 2017. PMID: 28091696
-
Lens dose reduction with a bismuth shield in neuro cone-beam computed tomography: an investigation on optimum shield device placement conditions.Radiol Phys Technol. 2022 Mar;15(1):25-36. doi: 10.1007/s12194-021-00644-0. Epub 2021 Nov 18. Radiol Phys Technol. 2022. PMID: 34796447
-
Breast dose reduction with organ-based, wide-angle tube current modulated CT.J Med Imaging (Bellingham). 2017 Jul;4(3):031208. doi: 10.1117/1.JMI.4.3.031208. Epub 2017 Aug 4. J Med Imaging (Bellingham). 2017. PMID: 28804730 Free PMC article.
-
[Dose optimization in CT examination of children].Radiologe. 2012 Oct;52(10):927-33. doi: 10.1007/s00117-012-2341-0. Radiologe. 2012. PMID: 22986576 German.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources