Reliability of shoulder range of motion comparing a goniometer to a digital level
- PMID: 20557263
- DOI: 10.3109/09593980903094230
Reliability of shoulder range of motion comparing a goniometer to a digital level
Abstract
The clinical use of digital levels, for joint measurement, may be a viable alternative to standard goniometry. The purpose of this study was to determine the intra- and intertester reliability of a construction grade digital level compared to the standard universal goniometer for measurements for active assisted shoulder range of motion (ROM). Two experienced physical therapists measured shoulder flexion, external rotation (ER), and internal rotation (IR) ROM bilaterally, on two different occasions, in 20 patients (9 males, 11 females, 18-79 years old) with unilateral shoulder pathology, using a goniometer and a digital level. Relative reliability was assessed by using intraclass correlation coefficients (ICC), and absolute reliability was assessed by using 95% limits of agreement (LOA). Intratester ICCs ranged from 0.91 to 0.99, and LOA ranged from 3 degrees to 9 degrees for measurements made with the goniometer and digital level. Intertester ICCs ranged from 0.31 to 0.95, and LOA ranged from 6 degrees to 25 degrees . For the comparison of goniometric vs. digital level ROM, ICCs ranged from 0.71 to 0.98. ER and IR ROM were 3-5 degrees greater for the digital level than the goniometer (p < 0.01). Goniometric vs. digital level LOA ranged from 6 degrees to 11 degrees for shoulder flexion. Both measurement techniques had excellent intratester reliability, but for intertester reliability ICCs were 20% lower and LOA were 2.3 times higher than intratester values. Reliability estimates were similar between the digital level and the goniometer. However, because glenohumeral rotation was 3-5 degrees greater for the digital level than the goniometer (systematic error), the two methods cannot be used interchangeably. On the basis of the average intratester LOA for the goniometer and the digital level, a change of 6-11 degrees is needed to be certain that true change has occurred. For comparison of measures made by two different therapists, a change is of 15 degrees is required to be certain a true change has occurred. A digital level can be used to reliably measure shoulder ROM but should not be used interchangeably with a standard goniometer.
Similar articles
-
The reliability and concurrent validity of scapular plane shoulder elevation measurements using a digital inclinometer and goniometer.Physiother Theory Pract. 2012 Feb;28(2):161-8. doi: 10.3109/09593985.2011.574203. Epub 2011 Jul 3. Physiother Theory Pract. 2012. PMID: 21721999
-
The reliability and minimal detectable change of shoulder mobility measurements using a digital inclinometer.Physiother Theory Pract. 2011 Feb;27(2):176-84. doi: 10.3109/09593985.2010.481011. Epub 2010 Aug 8. Physiother Theory Pract. 2011. PMID: 20690872
-
Active movement measurements of the shoulder girdle in healthy subjects with goniometer and tape measure techniques: a study on reliability and validity.Physiother Theory Pract. 2007 May-Jun;23(3):179-87. doi: 10.1080/09593980701209246. Physiother Theory Pract. 2007. PMID: 17558881
-
Intratester and intertester reliability and criterion validity of the parallelogram and universal goniometers for active knee flexion in healthy subjects.Physiother Res Int. 1997;2(3):150-66. doi: 10.1002/pri.97. Physiother Res Int. 1997. PMID: 9421820 Review.
-
The reliability and validity of goniometric elbow measurements in adults: A systematic review of the literature.Shoulder Elbow. 2018 Oct;10(4):274-284. doi: 10.1177/1758573218774326. Epub 2018 Jun 3. Shoulder Elbow. 2018. PMID: 30214494 Free PMC article. Review.
Cited by
-
Rehabilitation of Post-COVID-19 Musculoskeletal Sequelae in Geriatric Patients: A Case Series Study.Int J Environ Res Public Health. 2022 Nov 21;19(22):15350. doi: 10.3390/ijerph192215350. Int J Environ Res Public Health. 2022. PMID: 36430069 Free PMC article.
-
Canoe polo Athletes' Anthropometric, Physical, Nutritional, and Functional Characteristics and Performance in a Rowing Task: Cross-Sectional Study.Int J Environ Res Public Health. 2022 Oct 19;19(20):13518. doi: 10.3390/ijerph192013518. Int J Environ Res Public Health. 2022. PMID: 36294099 Free PMC article.
-
The Effect of Blood Flow Restriction Therapy on Shoulder Function Following Shoulder Stabilization Surgery: A Case Series.Int J Sports Phys Ther. 2022 Oct 2;17(6):1144-1155. doi: 10.26603/001c.37865. eCollection 2022. Int J Sports Phys Ther. 2022. PMID: 36873568 Free PMC article.
-
Postoperative Rehabilitation Protocol Following Arthroscopic Rotator Cuff Repair: A Prospective Single-Arm Pragmatic Interventional Study.Medicina (Kaunas). 2022 May 28;58(6):729. doi: 10.3390/medicina58060729. Medicina (Kaunas). 2022. PMID: 35743992 Free PMC article.
-
A webcam-based machine learning approach for three-dimensional range of motion evaluation.PLoS One. 2023 Oct 23;18(10):e0293178. doi: 10.1371/journal.pone.0293178. eCollection 2023. PLoS One. 2023. PMID: 37871043 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical