Helical tomotherapy for single and multiple liver tumours
- PMID: 20576108
- PMCID: PMC2900282
- DOI: 10.1186/1748-717X-5-58
Helical tomotherapy for single and multiple liver tumours
Abstract
Purpose: Dosimetric evaluations of single and multiple liver tumours performed using intensity-modulated helical tomotherapy (HT) were quantitatively investigated. Step-and-shoot intensity-modulated radiotherapy (SaS-IMRT) was used as a benchmark.
Methods: Sixteen patients separated into two groups with primary hepatocellular carcinomas or metastatic liver tumours previously treated using SaS-IMRT were examined and re-planned by HT. The dosimetric indices used included the conformity index (CI) and homogeneity index (HI) for the planned target volume (PTV), max/mean dose, quality index (QI), normal tissue complication probability (NTCP), V(30 Gy), and V(50%) for the specified organs at risk (OARs). The monitor units per fraction (MU/fr) and delivery time were also analysed.
Results: For the single tumour group, both planning systems satisfied the required PTV prescription, but no statistical significance was shown by the indexes checking. A shorter delivery time and lower MU/fr value were achieved by the SaS-IMRT. For the group of multiple tumours, the average improvement in CI and HI was 14% and 4% for HT versus SaS-IMRT, respectively. Lower V(50%), V(30 Gy) and QI values were found, indicating a significant dosimetric gain in HT. The NTCP value of the normal liver was 20.27 +/- 13.29% for SaS-IMRT and 2.38 +/- 2.25% for HT, indicating fewer tissue complications following HT. The latter also required a shorter delivery time.
Conclusions: Our study suggests dosimetric benefits of HT over SaS-IMRT plans in the case of multiple liver tumours, especially with regards sparing of OARs. No significant dosimetric difference was revealed in the case of single liver tumour, but SaS-IMRT showed better efficiency in terms of MU/fr and delivery time.
Figures


Similar articles
-
Dosimetric comparisons of helical tomotherapy and step-and-shoot intensity-modulated radiotherapy in nasopharyngeal carcinoma.Radiother Oncol. 2008 Oct;89(1):89-96. doi: 10.1016/j.radonc.2008.05.010. Epub 2008 Jun 2. Radiother Oncol. 2008. PMID: 18524401
-
Clinical dosimetric study of three radiotherapy techniques for postoperative breast cancer: Helical Tomotherapy, IMRT, and 3D-CRT.Technol Cancer Res Treat. 2011 Feb;10(1):15-23. doi: 10.7785/tcrt.2012.500174. Technol Cancer Res Treat. 2011. PMID: 21214284
-
Volumetric modulated arc therapy for nasopharyngeal carcinoma: a dosimetric comparison with TomoTherapy and step-and-shoot IMRT.Radiother Oncol. 2012 Sep;104(3):324-30. doi: 10.1016/j.radonc.2011.11.017. Epub 2012 Jan 9. Radiother Oncol. 2012. PMID: 22236614
-
Volumetric intensity-modulated Arc (RapidArc) therapy for primary hepatocellular carcinoma: comparison with intensity-modulated radiotherapy and 3-D conformal radiotherapy.Radiat Oncol. 2011 Jun 21;6:76. doi: 10.1186/1748-717X-6-76. Radiat Oncol. 2011. PMID: 21693003 Free PMC article.
-
Helical tomotherapy for head and neck squamous cell carcinoma: dosimetric comparison with linear accelerator-based step-and-shoot IMRT.J Cancer Res Ther. 2010 Apr-Jun;6(2):194-8. doi: 10.4103/0973-1482.65245. J Cancer Res Ther. 2010. PMID: 20622367
Cited by
-
Dose escalation by image-guided intensity-modulated radiotherapy leads to an increase in pain relief for spinal metastases: a comparison study with a regimen of 30 Gy in 10 fractions.Oncotarget. 2017 Jul 4;8(68):112330-112340. doi: 10.18632/oncotarget.18979. eCollection 2017 Dec 22. Oncotarget. 2017. PMID: 29348828 Free PMC article.
-
Dose verification in intensity modulation radiation therapy: a fractal dimension characteristics study.Biomed Res Int. 2013;2013:349437. doi: 10.1155/2013/349437. Epub 2013 Jul 8. Biomed Res Int. 2013. PMID: 23956976 Free PMC article.
-
Phase I dose escalation study of helical intensity-modulated radiotherapy-based stereotactic body radiotherapy for hepatocellular carcinoma.Oncotarget. 2016 Jun 28;7(26):40756-40766. doi: 10.18632/oncotarget.9450. Oncotarget. 2016. PMID: 27213593 Free PMC article. Clinical Trial.
-
Dosimetric advantages of generalised equivalent uniform dose-based optimisation on dose-volume objectives in intensity-modulated radiotherapy planning for bilateral breast cancer.Br J Radiol. 2012 Nov;85(1019):1499-506. doi: 10.1259/bjr/24112047. Br J Radiol. 2012. PMID: 23091290 Free PMC article.
-
Comparative analysis of SmartArc-based dual arc volumetric-modulated arc radiotherapy (VMAT) versus intensity-modulated radiotherapy (IMRT) for nasopharyngeal carcinoma.J Appl Clin Med Phys. 2011 Nov 15;12(4):3587. doi: 10.1120/jacmp.v12i4.3587. J Appl Clin Med Phys. 2011. PMID: 22089015 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials