Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2010 Jul;32(1):52-9.
doi: 10.1002/jmri.22223.

Visualization of cerebral microbleeds with dual-echo T2*-weighted magnetic resonance imaging at 7.0 T

Affiliations

Visualization of cerebral microbleeds with dual-echo T2*-weighted magnetic resonance imaging at 7.0 T

Mandy M A Conijn et al. J Magn Reson Imaging. 2010 Jul.

Abstract

Purpose: To assess the visualization of cerebral microbleeds with dual echo T2*-weighted imaging at 7.0 T magnetic resonance imaging (MRI).

Materials and methods: Ten consecutive participants (eight men, two women, mean age 54 +/- 12 years) with vascular disease or risk factors from the second manifestations of arterial disease (SMART) study were included. Dual-echo T2*-weighted scans (echo time: 2.5/15.0 msec) were made for all participants at 7.0 T MRI. The number of visible microbleeds and the diameter of the microbleeds were recorded on minimal intensity projection images of both echoes.

Results: : The first echo image shows dark microbleeds against a homogeneous, more hyperintense signal of the brain tissue without contrast for veins and basal ganglia. In eight patients microbleeds were observed, with a total of 104 microbleeds. Of these, 88 (84.6%) were visible on the first and 102 (98.0%) on the second echo. The mean diameter of the microbleeds was 1.24 mm for the first echo and 2.34 mm for the second echo.

Conclusion: T2*-weighted imaging at two echo times at 7.0 T combines the advantages of the first and second echo. Microbleeds visible on the first echo show large contrast with the surrounding tissue, even in the presence of paramagnetic ferritin. The second echo enables visualization of smaller microbleeds than the first echo.

PubMed Disclaimer

LinkOut - more resources