. 2011 Feb;39(2):185-9.
doi: 10.1016/j.ejcts.2010.05.020. Epub 2010 Jun 29.

The integrated place of tracheobronchial stents in the multidisciplinary management of large post-pneumonectomy fistulas: our experience using a novel customised conical self-expandable metallic stent

Affiliations

The integrated place of tracheobronchial stents in the multidisciplinary management of large post-pneumonectomy fistulas: our experience using a novel customised conical self-expandable metallic stent

Hervé Dutau et al. Eur J Cardiothorac Surg. 2011 Feb.

Abstract

Background: Stump dehiscence after pneumonectomy is a cause of morbidity and mortality in patients treated for non-small-cell lung carcinoma. Surgical repair remains the treatment of choice but can be postponed or contraindicated. Bronchoscopic techniques may be an option with curative intent or as a bridge towards definitive surgery. The aim of the study is to evaluate the efficacy and the outcome of a new customised covered conical self-expandable metallic stent in the management of large bronchopleural fistulas complicating pneumonectomies.

Methods: A case series using chart review of non-operable patients presenting with large bronchopleural fistulas (>6mm) post-pneumonectomies as a definitive treatment with curative intent for non-small-cell lung carcinomas and requiring the use of a dedicated conical shaped stent in two tertiary referral centres.

Results: Seven patients presenting large post-pneumonectomy fistulas (between 6 and 12 mm) were included. Cessation of the air leak and clinical improvement was achieved in all the patients after stent placement. Stent-related complications (two migrations and one stent rupture) were successfully managed using bronchoscopic techniques in two patients and surgery in one. Mortality, mainly related to overwhelming sepsis, was 57%. Delayed definitive surgery was achieved successfully in three patients (43%).

Conclusions: This case series assesses the short-term clinical efficacy of a new customised covered conical self-expandable metallic stent in the multidisciplinary management of large bronchopleural fistulas complicating pneumonectomies in patients deemed non-operable. Long-term benefits are jeopardised by infectious complications.