Rapid determination of C4-acylcarnitine and C5-acylcarnitine isomers in plasma and dried blood spots by UPLC-MS/MS as a second tier test following flow-injection MS/MS acylcarnitine profile analysis
- PMID: 20591710
- DOI: 10.1016/j.ymgme.2010.05.012
Rapid determination of C4-acylcarnitine and C5-acylcarnitine isomers in plasma and dried blood spots by UPLC-MS/MS as a second tier test following flow-injection MS/MS acylcarnitine profile analysis
Abstract
Background: Flow-injection MS/MS methods for elevated acylcarnitines are routinely performed in most newborn screening and biochemical genetics laboratories; however this technique cannot distinguish between isobaric compounds; therefore, chromatographic separation is required to quantitate isomers for differential diagnosis of some inborn errors of metabolism.
Methods: A UPLC-MS/MS method has been developed for the simultaneous quantitation of isobutyrylcarnitine and butyrylcarnitine, and a second UPLC-MS/MS method for the quantitation of isovalerylcarnitine, (S) and (R) 2-methylbutyrylcarnitine, pivaloylcarnitine and valerylcarnitine. Plasma and dried blood spots samples are extracted with methanol and derivatized with butanolic HCl. Deuterium labeled internal standards are used for quantitation. Separation is obtained using a methanol/water gradient with a C18 BEH, 1x100mm, 1.7microm UPLC column, at 60 degrees C; run time is less than 10min. The isomers are detected with a Quattro Premier triple quadrupole, with electrospray ionization in positive ion mode.
Results: Intra-day precision in plasma and dried blood spots ranged from 1.4% to 14% and accuracy from 88% to 114% respectively for butyrylcarnitine and isobutyrylcarnitine. Precision for the isomers of C5-acylcarnitine ranged from 1.3% to 15% and accuracy 87% to 119%, respectively in plasma or dried blood spots. Inter-day precision was within 20% at each concentration of isobutyrylcarnitine and butyrylcarnitine. Precision for 2-methylbutyrylcarnitine and isovalerylcarnitine at concentrations above the normal range was within 24%.
Conclusions: Two diagnostic tests based on the separation of C4-acylcarnitine and C5-acylcarnitine isomers by UPLC-MS/MS provide fast differential diagnosis of SCAD deficiency versus IBCD deficiency and IVA versus 2-MBCD deficiency. The separation of C5-acylcarnitines can reveal false elevation due to pivalic acid-containing antibiotics. Abnormal newborn screen results due to pivalate-generating prodrug antibiotics of maternal origin were confirmed. This separation of isomers can resolve multiple diagnostic challenges in both newborn screening and in cases with ambiguous metabolic test results.
Similar articles
-
A rapid UPLC-MS/MS method for simultaneous separation of 48 acylcarnitines in dried blood spots and plasma useful as a second-tier test for expanded newborn screening.Anal Bioanal Chem. 2012 Aug;404(3):741-51. doi: 10.1007/s00216-012-6194-1. Epub 2012 Jul 6. Anal Bioanal Chem. 2012. PMID: 22766757
-
UPLC-MS/MS analysis of C5-acylcarnitines in dried blood spots.Clin Chim Acta. 2013 Jun 5;421:41-5. doi: 10.1016/j.cca.2013.03.001. Epub 2013 Mar 13. Clin Chim Acta. 2013. PMID: 23499962
-
Measurement of free carnitine and acylcarnitines in plasma by HILIC-ESI-MS/MS without derivatization.J Chromatogr B Analyt Technol Biomed Life Sci. 2013 Aug 1;932:12-8. doi: 10.1016/j.jchromb.2013.05.028. Epub 2013 Jun 2. J Chromatogr B Analyt Technol Biomed Life Sci. 2013. PMID: 23816563
-
Application of mass spectrometry in newborn screening: about both small molecular diseases and lysosomal storage diseases.Top Curr Chem. 2014;336:177-96. doi: 10.1007/128_2012_354. Top Curr Chem. 2014. PMID: 22911488 Review.
-
Profiling of acylcarnitines and sterols from dried blood or plasma spot by atmospheric pressure thermal desorption chemical ionization (APTDCI) tandem mass spectrometry.Biochim Biophys Acta. 2011 Nov;1811(11):669-79. doi: 10.1016/j.bbalip.2011.05.009. Epub 2011 Jun 12. Biochim Biophys Acta. 2011. PMID: 21683155 Review.
Cited by
-
Emerging trends in paper spray mass spectrometry: Microsampling, storage, direct analysis, and applications.Mass Spectrom Rev. 2020 Jul;39(4):336-370. doi: 10.1002/mas.21601. Epub 2019 Sep 6. Mass Spectrom Rev. 2020. PMID: 31491055 Free PMC article. Review.
-
High-Protein Intake during Weight Loss Therapy Eliminates the Weight-Loss-Induced Improvement in Insulin Action in Obese Postmenopausal Women.Cell Rep. 2016 Oct 11;17(3):849-861. doi: 10.1016/j.celrep.2016.09.047. Cell Rep. 2016. PMID: 27732859 Free PMC article. Clinical Trial.
-
Validation of a dual LC-HRMS platform for clinical metabolic diagnosis in serum, bridging quantitative analysis and untargeted metabolomics.Metabolomics. 2014 Apr 4;10(2):312-323. doi: 10.1007/s11306-013-0582-1. Metabolomics. 2014. PMID: 25411574 Free PMC article.
-
Identification of urinary metabolites correlated with tacrolimus levels through high-precision liquid chromatography-mass spectrometry and machine learning algorithms in kidney transplant patients.Med Pharm Rep. 2025 Jan;98(1):125-134. doi: 10.15386/mpr-2805. Epub 2025 Jan 31. Med Pharm Rep. 2025. PMID: 39949902 Free PMC article.
-
Portuguese Neonatal Screening Program: A Cohort Study of 18 Years Using MS/MS.Int J Neonatal Screen. 2024 Mar 20;10(1):25. doi: 10.3390/ijns10010025. Int J Neonatal Screen. 2024. PMID: 38535129 Free PMC article.