. 2010 Sep;48(9):3117-21.
doi: 10.1128/JCM.00617-10. Epub 2010 Jun 30.

Screening of urine samples by flow cytometry reduces the need for culture

Affiliations

Screening of urine samples by flow cytometry reduces the need for culture

Santra Jolkkonen et al. J Clin Microbiol. 2010 Sep.

Abstract

Urine samples constitute a large proportion of samples tested in clinical microbiology laboratories. Culturing of the samples is fairly time- and labor-consuming, and most of the samples will yield no growth or insignificant growth. We analyzed the feasibility of the flow cytometry-based UF-500i instrument (Sysmex, Japan) to screen out urine samples with no growth or insignificant growth and reduce the number of samples to be cultured. A total of 1,094 urine specimens sent to our laboratory for culture during 4 months in the spring of 2009 in Lahti, Finland, were included in the study. After culture, all samples were analyzed with the Sysmex UF-500i for bacterial and leukocyte (white blood cell [WBC]) counts. Youden index and closest (0,1) methods were used to determine the cutoff values for bacterial and WBC counts in culture-positive and -negative groups. By flow cytometry, samples considered positive for UTI in culture had bacterial and WBC values that were significantly higher than those for samples considered negative. The flow cytometric screening worked best when both bacterial counts and WBC counts were used with age- and gender-specific cutoff values for all patient groups, excluding patients with urological disease or anomaly. By use of these cutoff values, 5/167 (3.0%) of culture-positive samples were missed by UF-500i and the percentage of samples that did not need to be cultured was 64.5%. Use of the UF-500i instrument is a reliable method for screening out a major part of the UTI-negative samples, significantly diminishing the amount of work required in the microbiology laboratory.

PubMed Disclaimer

References

    1. Akobeng, A. K. 2007. Understanding diagnostic tests 3: receiver operating characteristic curves. Acta Paediatr. 96:644-647. - PubMed
    1. Ben-Ezra, J., L. Bork, and R. McPherson. 1998. Evaluation of the Sysmex UF-100 automated urinalysis analyzer. Clin. Chem. 44:92-95. - PubMed
    1. Brilha, S., H. Proença, J. M. Cristino, and T. Hänscheid. 2010. Use of flow cytometry (Sysmex® UF-100) to screen for positive urine cultures: in search for ideal cut-off. Clin. Chem. Lab. Med. 48:289-292. - PubMed
    1. Devillé, W. L. J. M., J. C. Yzermans, N. P. van Duijn, P. D. Bezemer, D. A. W. M. van der Windt, and L. M. Bouter. 2004. The urine dipstick test useful to rule out infections. A meta-analysis of the accuracy. BMC Urol. 4:4. - PMC - PubMed
    1. dos Santos, J. C., L. P. Weber, and L. R. R. Perez. 2007. Evaluation of urinalysis parameters to predict urinary-tract infection. Braz. J. Infect. Dis. 11:479-481. - PubMed

Publication types