Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2010 Aug;2(8):751-6.
doi: 10.1016/j.pmrj.2010.01.016. Epub 2010 Jul 3.

Incidence and identification of intrathecal baclofen catheter malfunction

Affiliations

Incidence and identification of intrathecal baclofen catheter malfunction

Eric M Dvorak et al. PM R. 2010 Aug.

Abstract

Objective: A retrospective chart review was undertaken of cases of intrathecal baclofen (ITB) pump/catheter malfunction and the diagnostic tests performed to identify the problem. An internal review was performed to develop a diagnostic flow chart to have a systematic method for identifying ITB pump and catheter complications.

Design: Retrospective chart review.

Setting: Tertiary care hospital.

Participants: A total of 167 adult patients currently followed in outpatient clinic with intractable spasticity and ITB pump implanted between January 1994 and May 2009.

Interventions: None.

Main outcome measures: Catheter malfunction was identified either by anterior/posterior and lateral thoracic/lumbar radiographs, fluoroscopic guided side port access, contrast agent injection followed by CT (fluoro/CT) scan, or indium radionuclide studies.

Results: During the study period, 33 patients had 37 catheter revisions. Radiographs were obtained in all cases; fluoro/CT studies in 22, and indium studies in 6. Four cases had both fluoro/CT and indium studies. A total of 13 cases (35.1%) were diagnosed with radiographs; 9 cases (24.3%) were diagnosed by inability to withdraw cerebral spinal fluid from the side port; 13 cases (35.1%) were diagnosed with fluoro/CT studies; and 2 cases (5.4%) were diagnosed with indium studies. Fluoro/CT studies demonstrated subdural catheter location in 7 cases. A total of 2 of 4 cases with both fluoro/CT and indium studies had normal-appearing indium scans and an abnormal fluoro/CT study confirming subdural catheter placement.

Conclusions: On the basis of an internal review of the diagnostic studies used to identify patients with suspected ITB pump/catheter malfunction, a troubleshooting flow chart was developed. Timely identification and correction of potential ITB complications could improve the clinical effectiveness of ITB and may reduce unnecessary health-care costs.

PubMed Disclaimer

Similar articles

Cited by

MeSH terms

LinkOut - more resources