Continuous glucose monitoring in subcutaneous tissue using factory-calibrated sensors: a pilot study
- PMID: 20615099
- DOI: 10.1089/dia.2010.0051
Continuous glucose monitoring in subcutaneous tissue using factory-calibrated sensors: a pilot study
Abstract
Background: Commercial continuous subcutaneous glucose monitors require in vivo calibration using capillary blood glucose tests. Feasibility of factory calibration, i.e., sensor batch characterization in vitro with no further need for in vivo calibration, requires a predictable and stable in vivo sensor sensitivity and limited inter- and intra-subject variation of the ratio of interstitial to blood glucose concentration.
Methods: Twelve volunteers wore two FreeStyle Navigator (Abbott Diabetes Care, Alameda, CA) continuous glucose monitoring systems for 5 days in parallel for two consecutive sensor wears (four sensors per subject, 48 sensors total). Sensors from a prototype sensor lot with a low variability in glucose sensitivity were used for the study. Median sensor sensitivity values based on capillary blood glucose were calculated per sensor and compared for inter- and intra-subject variation. Mean absolute relative difference (MARD) calculation and error grid analysis were performed using a single calibration factor for all sensors to simulate factory calibration and compared to standard fingerstick calibration.
Results: Sensor sensitivity variation in vitro was 4.6%, which increased to 8.3% in vivo (P < 0.0001). Analysis of variance revealed no significant inter-subject differences in sensor sensitivity (P = 0.134). Applying a single universal calibration factor retrospectively to all sensors resulted in a MARD of 10.4% and 88.1% of values in Clarke Error Grid Zone A, compared to a MARD of 10.9% and 86% of values in Error Grid Zone A for fingerstick calibration.
Conclusions: Factory calibration of sensors for continuous subcutaneous glucose monitoring is feasible with similar accuracy to standard fingerstick calibration. Additional data are required to confirm this result in subjects with diabetes.
Similar articles
-
Feasibility of Factory Calibration for Subcutaneous Glucose Sensors in Subjects With Diabetes.J Diabetes Sci Technol. 2014 Jan;8(1):89-94. doi: 10.1177/1932296813511747. Epub 2014 Jan 1. J Diabetes Sci Technol. 2014. PMID: 24876543 Free PMC article.
-
Influence of time point of calibration on accuracy of continuous glucose monitoring in individuals with type 1 diabetes.Diabetes Technol Ther. 2012 Jul;14(7):583-8. doi: 10.1089/dia.2011.0271. Epub 2012 Apr 18. Diabetes Technol Ther. 2012. PMID: 22512266
-
Accuracy of FreeStyle Libre in Adults with Type 1 Diabetes: The Effect of Sensor Age.Diabetes Technol Ther. 2020 Mar;22(3):203-207. doi: 10.1089/dia.2019.0262. Epub 2020 Jan 23. Diabetes Technol Ther. 2020. PMID: 31613140 Clinical Trial.
-
Continuous glucose monitoring with glucose sensors: calibration and assessment criteria.Diabetes Technol Ther. 2003;5(4):572-86. doi: 10.1089/152091503322250596. Diabetes Technol Ther. 2003. PMID: 14511412 Review.
-
Amperometric glucose sensors: sources of error and potential benefit of redundancy.J Diabetes Sci Technol. 2010 Jan 1;4(1):221-5. doi: 10.1177/193229681000400127. J Diabetes Sci Technol. 2010. PMID: 20167187 Free PMC article. Review.
Cited by
-
Factory-Calibrated Continuous Glucose Sensors: The Science Behind the Technology.Diabetes Technol Ther. 2017 May;19(S2):S44-S50. doi: 10.1089/dia.2017.0025. Diabetes Technol Ther. 2017. PMID: 28541139 Free PMC article.
-
Predicting Plasma Glucose From Interstitial Glucose Observations Using Bayesian Methods.J Diabetes Sci Technol. 2014 Mar;8(2):321-330. doi: 10.1177/1932296814523878. Epub 2014 Mar 6. J Diabetes Sci Technol. 2014. PMID: 24876584 Free PMC article.
-
Estimating plasma glucose from interstitial glucose: the issue of calibration algorithms in commercial continuous glucose monitoring devices.Sensors (Basel). 2010;10(12):10936-52. doi: 10.3390/s101210936. Epub 2010 Dec 3. Sensors (Basel). 2010. PMID: 22163505 Free PMC article. Review.
-
Modulation of the foreign body reaction for implants in the subcutaneous space: microdialysis probes as localized drug delivery/sampling devices.J Diabetes Sci Technol. 2011 May 1;5(3):619-31. doi: 10.1177/193229681100500316. J Diabetes Sci Technol. 2011. PMID: 21722577 Free PMC article.
-
Feasibility of Factory Calibration for Subcutaneous Glucose Sensors in Subjects With Diabetes.J Diabetes Sci Technol. 2014 Jan;8(1):89-94. doi: 10.1177/1932296813511747. Epub 2014 Jan 1. J Diabetes Sci Technol. 2014. PMID: 24876543 Free PMC article.