A fast, angle-dependent, analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems
- PMID: 20632571
- PMCID: PMC2885940
- DOI: 10.1118/1.3397462
A fast, angle-dependent, analytical model of CsI detector response for optimization of 3D x-ray breast imaging systems
Erratum in
- Med Phys. 2011 Apr;38(4):2307
Abstract
Purpose: Accurate models of detector blur are crucial for performing meaningful optimizations of three-dimensional (3D) x-ray breast imaging systems as well as for developing reconstruction algorithms that faithfully reproduce the imaged object anatomy. So far, x-ray detector blur has either been ignored or modeled as a shift-invariant symmetric function for these applications. The recent development of a Monte Carlo simulation package called MANTIS has allowed detailed modeling of these detector blur functions and demonstrated the magnitude of the anisotropy for both tomosynthesis and breast CT imaging systems. Despite the detailed results that MANTIS produces, the long simulation times required make inclusion of these results impractical in rigorous optimization and reconstruction algorithms. As a result, there is a need for detector blur models that can be rapidly generated.
Methods: In this study, the authors have derived an analytical model for deterministic detector blur functions, referred to here as point response functions (PRFs), of columnar CsI phosphor screens. The analytical model is x-ray energy and incidence angle dependent and draws on results from MANTIS to indirectly include complicated interactions that are not explicitly included in the mathematical model. Once the mathematical expression is derived, values of the coefficients are determined by a two-dimensional (2D) fit to MANTIS-generated results based on a figure-of-merit (FOM) that measures the normalized differences between the MANTIS and analytical model results averaged over a region of interest. A smaller FOM indicates a better fit. This analysis was performed for a monochromatic x-ray energy of 25 keV, a CsI scintillator thickness of 150 microm, and four incidence angles (0 degrees, 15 degrees, 30 degrees, and 45 degrees).
Results: The FOMs comparing the analytical model to MANTIS for these parameters were 0.1951 +/- 0.0011, 0.1915 +/- 0.0014, 0.2266 +/- 0.0021, and 0.2416 +/- 0.0074 for 0 degrees, 15 degrees, 30 degrees, and 45 degrees, respectively. As a comparison, the same FOMs comparing MANTIS to 2D symmetric Gaussian fits to the zero-angle PRF were 0.6234 +/- 0.0020, 0.9058 +/- 0.0029, 1.491 +/- 0.012, and 2.757 +/- 0.039 for the same set of incidence angles. Therefore, the analytical model matches MANTIS results much better than a 2D symmetric Gaussian function. A comparison was also made against experimental data for a 170 microm thick CsI screen and an x-ray energy of 25.6 keV. The corresponding FOMs were 0.3457 +/- 0.0036, 0.3281 +/- 0.0057, 0.3422 +/- 0.0023, and 0.3677 +/- 0.0041 for 0 degrees, 15 degrees, 30 degrees, and 45 degrees, respectively. In a previous study, FOMs comparing the same experimental data to MANTIS PRFs were found to be 0.2944 +/- 0.0027, 0.2387 +/- 0.0039, 0.2816 +/- 0.0025, and 0.2665 +/- 0.0032 for the same set of incidence angles.
Conclusions: The two sets of derived FOMs, comparing MANTIS-generated PRFs and experimental data to the analytical model, demonstrate that the analytical model is able to reproduce experimental data with a FOM of less than two times that comparing MANTIs and experimental data. This performance is achieved in less than one millionth the computation time required to generate a comparable PRF with MANTIS. Such small computation times will allow for the inclusion of detailed detector physics in rigorous optimization and reconstruction algorithms for 3D x-ray breast imaging systems.
Figures











Similar articles
-
Experimental validation of Monte Carlo (MANTIS) simulated x-ray response of columnar CsI scintillator screens.Med Phys. 2009 Nov;36(11):4944-56. doi: 10.1118/1.3233683. Med Phys. 2009. PMID: 19994503 Free PMC article.
-
Anisotropic imaging performance in breast tomosynthesis.Med Phys. 2007 Nov;34(11):4076-91. doi: 10.1118/1.2779943. Med Phys. 2007. PMID: 18074617
-
Anisotropic imaging performance in indirect x-ray imaging detectors.Med Phys. 2006 Aug;33(8):2698-713. doi: 10.1118/1.2208925. Med Phys. 2006. PMID: 16967568
-
[Digital breast tomosynthesis : technical principles, current clinical relevance and future perspectives].Radiologe. 2010 Nov;50(11):991-8. doi: 10.1007/s00117-010-2008-7. Radiologe. 2010. PMID: 21052620 Review. German.
-
OsiriX: an open-source software for navigating in multidimensional DICOM images.J Digit Imaging. 2004 Sep;17(3):205-16. doi: 10.1007/s10278-004-1014-6. Epub 2004 Jun 29. J Digit Imaging. 2004. PMID: 15534753 Free PMC article. Review.
Cited by
-
Digital Breast Tomosynthesis: State of the Art.Radiology. 2015 Dec;277(3):663-84. doi: 10.1148/radiol.2015141303. Radiology. 2015. PMID: 26599926 Free PMC article.
-
Theoretical and Monte Carlo optimization of a stacked three-layer flat-panel x-ray imager for applications in multi-spectral medical imaging.Proc SPIE Int Soc Opt Eng. 2016;9783:97833Z. doi: 10.1117/12.2217085. Epub 2016 Mar 29. Proc SPIE Int Soc Opt Eng. 2016. PMID: 28845080 Free PMC article.
-
Proposing a New Velocity Profile for Continuous X-Ray Tube Motion in Digital Breast Tomosynthesis.Proc SPIE Int Soc Opt Eng. 2013 Feb;8668:86680F. doi: 10.1117/12.2007764. Epub 2013 Mar 19. Proc SPIE Int Soc Opt Eng. 2013. PMID: 38800605 Free PMC article.
-
Modelling the transport of optical photons in scintillation detectors for diagnostic and radiotherapy imaging.Phys Med Biol. 2017 Oct 4;62(20):R207-R235. doi: 10.1088/1361-6560/aa8b31. Phys Med Biol. 2017. PMID: 28976914 Free PMC article.
-
Non-stationary model of oblique x-ray incidence in amorphous selenium detectors: I. Point spread function.Med Phys. 2019 Feb;46(2):494-504. doi: 10.1002/mp.13313. Epub 2019 Jan 16. Med Phys. 2019. PMID: 30488462 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials