Comparison of automated oscillometric versus auscultatory blood pressure measurement
- PMID: 20643251
- DOI: 10.1016/j.amjcard.2010.03.040
Comparison of automated oscillometric versus auscultatory blood pressure measurement
Abstract
Most clinical offices rely on automated oscillometric devices to measure blood pressure (BP), but the accuracy of this technique versus auscultatory determination using a mercury manometer is controversial. To assess the accuracy of automated oscillometric readings, BP was measured from the same site and cuff, in 337 consecutive patients seen in a routine cardiology office, using a simultaneous connection to an automated oscillometric and a mercury manometer technique. The mean systolic BP (133 +/- 20 mm Hg) and diastolic BP (72 +/- 11 mm Hg) were significantly greater using the mercury manometer than the automated oscillometric technique (systolic 131 +/- 18 and diastolic 70 +/- 12 mm Hg, p <0.0001). Discrepancies (almost always lower oscillometric and greater mercury manometer) in systolic BP were seen in 22% of all patients. Discrepancies in diastolic BP were seen in 20% of all patients. The mean of the discrepancy between the 2 techniques was 1.95 +/- 5 mm Hg (range 1 to 26) for systolic BP and 1.3 +/- 4 mm Hg (range 1 to 25) for diastolic BP. The discrepancies were greater in patients >65 years. In conclusion, the mercury manometer technique resulted in consistently greater BP values than oscillometric devices. These findings have important clinical implications, including the concept that patients whose BP appears to be under control using the oscillometric technique might not be at their goal BP and might have been undertreated.
Copyright (c) 2010 Elsevier Inc. All rights reserved.
Comment in
-
Oscillometric blood pressure values are algorithm-specific.Am J Cardiol. 2010 Nov 15;106(10):1524; author reply 1524-5. doi: 10.1016/j.amjcard.2010.08.030. Am J Cardiol. 2010. PMID: 21059454 No abstract available.
Similar articles
-
Discarding the first of three nurse-auscultatory or oscillometric blood pressure measurements does not improve the association of office blood pressure with ABPM.Blood Press Monit. 2010 Jun;15(3):146-51. doi: 10.1097/MBP.0b013e328337ce76. Blood Press Monit. 2010. PMID: 20407368
-
Effect of the shapes of the oscillometric pulse amplitude envelopes and their characteristic ratios on the differences between auscultatory and oscillometric blood pressure measurements.Blood Press Monit. 2007 Oct;12(5):297-305. doi: 10.1097/MBP.0b013e32826fb773. Blood Press Monit. 2007. PMID: 17890968
-
Replacing the mercury manometer with an oscillometric device in a hypertension clinic: implications for clinical decision making.J Hum Hypertens. 2011 Nov;25(11):692-8. doi: 10.1038/jhh.2010.107. Epub 2010 Nov 25. J Hum Hypertens. 2011. PMID: 21107434
-
Oscillometric blood pressure: a review for clinicians.J Am Soc Hypertens. 2014 Dec;8(12):930-8. doi: 10.1016/j.jash.2014.08.014. Epub 2014 Sep 2. J Am Soc Hypertens. 2014. PMID: 25492837 Review.
-
Automated 'oscillometric' blood pressure measuring devices: how they work and what they measure.J Hum Hypertens. 2023 Feb;37(2):93-100. doi: 10.1038/s41371-022-00693-x. Epub 2022 May 30. J Hum Hypertens. 2023. PMID: 35637256 Free PMC article. Review.
Cited by
-
Paediatric Hypertension in Africa: A Systematic Review and Meta-Analysis.EClinicalMedicine. 2021 Dec 6;43:101229. doi: 10.1016/j.eclinm.2021.101229. eCollection 2022 Jan. EClinicalMedicine. 2021. PMID: 34917909 Free PMC article.
-
Estimation of Physiologic Pressures: Invasive and Non-Invasive Techniques, AI Models, and Future Perspectives.Sensors (Basel). 2023 Jun 20;23(12):5744. doi: 10.3390/s23125744. Sensors (Basel). 2023. PMID: 37420919 Free PMC article. Review.
-
Effectiveness of in-office blood pressure measurement by eye care practitioners in early detection and management of hypertension.Int J Ophthalmol. 2015 Jun 18;8(3):612-21. doi: 10.3980/j.issn.2222-3959.2015.03.32. eCollection 2015. Int J Ophthalmol. 2015. PMID: 26086018 Free PMC article.
-
Improved Measurement of Blood Pressure by Extraction of Characteristic Features from the Cuff Oscillometric Waveform.Sensors (Basel). 2015 Jun 16;15(6):14142-61. doi: 10.3390/s150614142. Sensors (Basel). 2015. PMID: 26087370 Free PMC article.
-
Effects of Resistance Exercise with and without Blood Flow Restriction on Acute Hemodynamic Responses: A Systematic Review and Meta-Analysis.Life (Basel). 2024 Jun 28;14(7):826. doi: 10.3390/life14070826. Life (Basel). 2024. PMID: 39063580 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources