Computer-aided detection; the effect of training databases on detection of subtle breast masses
- PMID: 20650667
- PMCID: PMC2952663
- DOI: 10.1016/j.acra.2010.06.009
Computer-aided detection; the effect of training databases on detection of subtle breast masses
Abstract
Rationale and objectives: Lesion conspicuity is typically highly correlated with visual difficulty for lesion detection, and computer-aided detection (CAD) has been widely used as a "second reader" in mammography. Hence, increasing CAD sensitivity in detecting subtle cancers without increasing false-positive rates is important. The aim of this study was to investigate the effect of training database case selection on CAD performance in detecting low-conspicuity breast masses.
Materials and methods: A full-field digital mammographic image database that included 525 cases depicting malignant masses was randomly partitioned into three subsets. A CAD scheme was applied to detect all initially suspected mass regions and compute region conspicuity. Training samples were iteratively selected from two of the subsets. Four types of training data sets-(1) one including all available true-positive mass regions in the two subsets ("all"), (2) one including 350 randomly selected mass regions ("diverse"), (3) one including 350 high-conspicuity mass regions ("easy"), and (4) one including 350 low-conspicuity mass regions ("difficult")-were assembled. In each training data set, the same number of randomly selected false-positive regions as the true-positives were also included. Two classifiers, an artificial neural network (ANN) and a k-nearest neighbor (KNN) algorithm, were trained using each of the four training data sets and tested on all suspected regions in the remaining data set. Using a threefold cross-validation method, the performance changes of the CAD schemes trained using one of the four training data sets were computed and compared.
Results: CAD initially detected 1025 true-positive mass regions depicted on 507 cases (97% case-based sensitivity) and 9569 false-positive regions (3.5 per image) in the entire database. Using the all training data set, CAD achieved the highest overall performance on the entire testing database. However, CAD detected the highest number of low-conspicuity masses when the difficult training data set was used for training. Results did agree for both ANN-based and KNN-based classifiers in all tests. Compared to the use of the all training data set, the sensitivity of the schemes trained using the difficult data set decreased by 8.6% and 8.4% for the ANN and KNN algorithm on the entire database, respectively, but the detection of low-conspicuity masses increased by 7.1% and 15.1% for the ANN and KNN algorithm at a false-positive rate of 0.3 per image.
Conclusions: CAD performance depends on the size, diversity, and difficulty level of the training database. To increase CAD sensitivity in detecting subtle cancer, one should increase the fraction of difficult cases in the training database rather than simply increasing the training data set size.
Copyright © 2010 AUR. Published by Elsevier Inc. All rights reserved.
Figures


Similar articles
-
Improving performance of computer-aided detection scheme by combining results from two machine learning classifiers.Acad Radiol. 2009 Mar;16(3):266-74. doi: 10.1016/j.acra.2008.08.012. Acad Radiol. 2009. PMID: 19201355 Free PMC article.
-
A method to test the reproducibility and to improve performance of computer-aided detection schemes for digitized mammograms.Med Phys. 2004 Nov;31(11):2964-72. doi: 10.1118/1.1806291. Med Phys. 2004. PMID: 15587648
-
Dual system approach to computer-aided detection of breast masses on mammograms.Med Phys. 2006 Nov;33(11):4157-68. doi: 10.1118/1.2357838. Med Phys. 2006. PMID: 17153394 Free PMC article.
-
Needs assessment for next generation computer-aided mammography reference image databases and evaluation studies.Int J Comput Assist Radiol Surg. 2011 Nov;6(6):749-67. doi: 10.1007/s11548-011-0553-9. Epub 2011 Mar 30. Int J Comput Assist Radiol Surg. 2011. PMID: 21448711 Review.
-
A review of automatic mass detection and segmentation in mammographic images.Med Image Anal. 2010 Apr;14(2):87-110. doi: 10.1016/j.media.2009.12.005. Epub 2009 Dec 29. Med Image Anal. 2010. PMID: 20071209 Review.
Cited by
-
A new and fast image feature selection method for developing an optimal mammographic mass detection scheme.Med Phys. 2014 Aug;41(8):081906. doi: 10.1118/1.4890080. Med Phys. 2014. PMID: 25086537 Free PMC article.
-
Assessing the predictive accuracy of lung cancer, metastases, and benign lesions using an artificial intelligence-driven computer aided diagnosis system.Quant Imaging Med Surg. 2021 Aug;11(8):3629-3642. doi: 10.21037/qims-20-1314. Quant Imaging Med Surg. 2021. PMID: 34341737 Free PMC article.
-
Applying a new computer-aided detection scheme generated imaging marker to predict short-term breast cancer risk.Phys Med Biol. 2018 May 15;63(10):105005. doi: 10.1088/1361-6560/aabefe. Phys Med Biol. 2018. PMID: 29667606 Free PMC article.
-
Optimization of Network Topology in Computer-Aided Detection Schemes Using Phased Searching with NEAT in a Time-Scaled Framework.Cancer Inform. 2014 Oct 13;13(Suppl 1):17-27. doi: 10.4137/CIN.S13885. eCollection 2014. Cancer Inform. 2014. PMID: 25392680 Free PMC article. Review.
-
Applying a new quantitative image analysis scheme based on global mammographic features to assist diagnosis of breast cancer.Comput Methods Programs Biomed. 2019 Oct;179:104995. doi: 10.1016/j.cmpb.2019.104995. Epub 2019 Jul 29. Comput Methods Programs Biomed. 2019. PMID: 31443864 Free PMC article.
References
-
- Nishikawa RM, Kallergi M. Computer-aided detection, in its present form, is not an effective aid for screening mammography, (Point/Counterpoint) Med Phys. 2006;33:811–814. - PubMed
-
- Freer TM, Ulissey MJ. Screening mammography with computer-aided detection: prospective study of 12,860 patients in a community breast center. Radiology. 2001;220:781–786. - PubMed
-
- Gur D, Sumkin JH, Rockette HE, et al. Changes in breast cancer detection and mammography recall rates after the introduction of a computer-aided detection system. J Natl Cancer Inst. 2004;96:185–190. - PubMed
-
- Khoo LA, Taylor P, Given-Wilson RM. Computer-aided detection in the United Kingdom National Breast Screening Programme: prospective study. Radiology. 2005;237:444–449. - PubMed
-
- Morton MJ, Whaley DH, Brandt KR, Amrami KK. Screening mammograms: Interpretation with computer-aided detection – prospective evaluation. Radiology. 2006;239:375–383. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous