Evaluation of intraperitoneal placement of absorbable and nonabsorbable barrier coated mesh secured with fibrin sealant in a New Zealand white rabbit model
- PMID: 20652323
- DOI: 10.1007/s00464-010-1230-8
Evaluation of intraperitoneal placement of absorbable and nonabsorbable barrier coated mesh secured with fibrin sealant in a New Zealand white rabbit model
Abstract
Background: This study aimed to evaluate the acute and chronic fixation strength of fibrin sealant (FS) as an alternative method of fixation for laparoscopic ventral hernia repair (LVHR).
Methods: Representative mesh types for LVHR included one nonabsorbable barrier mesh (Composix) and three absorbable barrier meshes (Sepramesh, Proceed, and Parietex composite). Macroporous polypropylene mesh (Prolite Ultra) served as the control mesh. Three methods of fixation were used, namely, 0-polypropylene suture+FS (ARTISS 4 IU), FS alone (ARTISS), and tacks alone, to secure 3×4-cm pieces of mesh (10 of each combination) to the peritoneal surface of New Zealand white rabbit abdominal wall. After 2 h of incubation at 37 °C, specimens underwent acute testing. Subsequently, a chronic phase was completed using the aforementioned fixation methods (10 of each combination), in which two 4×4-cm pieces of mesh were secured intraperitoneally in each of 75 New Zealand white rabbits, which survived 8 weeks until they were sacrificed. A transparent grid overlay was used to measure the mesh and adhesion area. Adhesion tenacity was characterized using the Garrard adhesion scale. In both the acute and chronic samples, a 3×3-cm area of mesh-tissue interface underwent lap shear testing at a rate of 0.42 mm/s using a tensiometer (Instron 5542). The maximum load sustained by the mesh-tissue construct was recorded as the acute fixation strength in newtons (N). Data are given as means±standard error of the mean. Statistical significance (p<0.05) was determined using a one-way analysis of variance (ANOVA) with Fisher's least significant difference (LSD) posttest or a nonparametric Kruskal-Wallis test (adhesion scores).
Results: The acute fixation strength was significantly greater for all the meshes secured with either suture+FS or tacks alone than for FS alone (p<0.001 for all comparisons). All the meshes except Proceed demonstrated greater acute fixation strength with suture+FS than with tacks alone (p≤0.016). Composix achieved greater acute fixation with suture+FS than all the other meshes (p≤0.022). Acute fixation with suture + FS was greater for Parietex Composite and ProLite Ultra than for Proceed (p≤0.015). When the animals were sacrificed, 48 of 50 meshes fixed with FS alone were insufficiently affixed to the abdominal wall, which may have resulted in hernia recurrence in a hernia model. The chronic fixation strength was greater for all the mesh types with either suture+FS or tacks only than with FS alone (p≤0.0005). The chronic fixation strength was greater with suture+FS than with tacks for Proceed and ProLite Ultra (p≤0.013). Neither mesh area nor adhesion tenacity differed significantly with any mesh/fixation method combination.
Conclusions: In a chronic rabbit model of LVHR, fixation strength with FS alone was inadequate for selected nonabsorbable and absorbable barrier-coated meshes. The acute and chronic fixation strengths of suture+FS were equivalent or superior to the fixation strength of tacks alone. Using a combination of suture and FS for mesh fixation in LVHR may provide adequate fixation while decreasing postoperative pain due to spiral titanium tacks. In this preclinical series, mesh secured to the peritoneal surface by FS alone may have led to early recurrence.
Similar articles
-
Evaluation of acute fixation strength of absorbable and nonabsorbable barrier coated mesh secured with fibrin sealant.Hernia. 2010 Oct;14(5):505-9. doi: 10.1007/s10029-010-0670-3. Epub 2010 May 9. Hernia. 2010. PMID: 20454989
-
Histologic evaluation of absorbable and non-absorbable barrier coated mesh secured to the peritoneum with fibrin sealant in a New Zealand white rabbit model.Hernia. 2011 Dec;15(6):677-84. doi: 10.1007/s10029-011-0834-9. Epub 2011 May 24. Hernia. 2011. PMID: 21607571 Free PMC article.
-
Evaluation of adhesion formation, mesh fixation strength, and hydroxyproline content after intraabdominal placement of polytetrafluoroethylene mesh secured using titanium spiral tacks, nitinol anchors, and polypropylene suture or polyglactin 910 suture.Surg Endosc. 2005 Jun;19(6):780-5. doi: 10.1007/s00464-004-8927-5. Epub 2005 Mar 23. Surg Endosc. 2005. PMID: 15776210
-
Mesh fixation techniques in primary ventral or incisional hernia repair.Cochrane Database Syst Rev. 2021 May 28;5(5):CD011563. doi: 10.1002/14651858.CD011563.pub2. Cochrane Database Syst Rev. 2021. PMID: 34046884 Free PMC article.
-
Pain and Surgical Outcomes Reporting After Laparoscopic Ventral Hernia Repair in Relation to Mesh Fixation Technique: A Systematic Review and Meta-Analysis of Randomized Clinical Trials.J Laparoendosc Adv Surg Tech A. 2018 Nov;28(11):1298-1315. doi: 10.1089/lap.2017.0609. Epub 2018 Jun 18. J Laparoendosc Adv Surg Tech A. 2018. PMID: 29912610
Cited by
-
Remodeling characteristics and biomechanical properties of a crosslinked versus a non-crosslinked porcine dermis scaffolds in a porcine model of ventral hernia repair.Hernia. 2015 Apr;19(2):207-18. doi: 10.1007/s10029-013-1070-2. Epub 2013 Mar 13. Hernia. 2015. PMID: 23483265 Free PMC article.
-
Developing a Hernia Mesh Tissue Integration Index Using a Porcine Model-A Pilot Study.Front Surg. 2020 Nov 26;7:600195. doi: 10.3389/fsurg.2020.600195. eCollection 2020. Front Surg. 2020. PMID: 33324674 Free PMC article.
-
Early biocompatibility of crosslinked and non-crosslinked biologic meshes in a porcine model of ventral hernia repair.Hernia. 2011 Apr;15(2):157-64. doi: 10.1007/s10029-010-0770-0. Epub 2011 Jan 8. Hernia. 2011. PMID: 21222009 Free PMC article.
-
Remodeling characteristics and collagen distribution in biological scaffold materials explanted from human subjects after abdominal soft tissue reconstruction: an analysis of scaffold remodeling characteristics by patient risk factors and surgical site classifications.Ann Surg. 2015 Feb;261(2):405-15. doi: 10.1097/SLA.0000000000000471. Ann Surg. 2015. PMID: 24374547 Free PMC article. Clinical Trial.
-
In vivo degradation of 14C-labeled porcine dermis biologic scaffold.Biomaterials. 2014 Sep;35(29):8297-304. doi: 10.1016/j.biomaterials.2014.06.015. Epub 2014 Jul 3. Biomaterials. 2014. PMID: 24997479 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials