Effect of 3 excimer laser ablation frequencies (200 Hz, 500 Hz, 1000 Hz) on the cornea using a 1000 Hz scanning-spot excimer laser
- PMID: 20656164
- DOI: 10.1016/j.jcrs.2010.01.033
Effect of 3 excimer laser ablation frequencies (200 Hz, 500 Hz, 1000 Hz) on the cornea using a 1000 Hz scanning-spot excimer laser
Abstract
Purpose: To evaluate the possible side effects on human and porcine cadaver eyes induced by excimer laser ablation with 3 ablation frequencies.
Setting: Augenklinik, Klinikum rechts der Isar, Technische Universitaet Muenchen, Munich, Germany.
Methods: Central epithelial abrasion was performed on porcine and human donor eyes. Next, the porcine eyes were photoablated (-9.00 diopters) at 1 of 3 frequencies (200 Hz, 500 Hz, or 1000 Hz) using a prototype 1000 Hz excimer laser. The human eyes were ablated at 1000 Hz. The surface of the stroma as well as the structure and ultrastructure of the corneal cells and stroma were examined using light microscopy, transmission electron microscopy, and scanning electron microscopy (SEM). Special attention was given to the detection of potential thermal damage and the evaluation of corneal smoothness.
Results: Histopathologic examination showed very low to almost no differences between the 3 repetition rates. In all cases, SEM showed an equally smooth surface.
Conclusions: The structural and ultrastructural evaluation of corneas treated with ablation frequencies of 200 Hz, 500 Hz, and 1000 Hz showed no specific side effects associated with the high repetition rates. The ablation quality was comparable in the 3 frequency groups. However, the treatment time was significantly less with a high repetition rate, indicating the clinical potential in refractive surgery of excimer lasers with a repetition rate of 1000 Hz.
Financial disclosure: No author has a financial or proprietary interest in any material or method mentioned. Additional disclosures are found in the footnotes.
Copyright 2010 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Effects of laser repetition rate on corneal tissue ablation for 193-nm excimer laser light.Lasers Surg Med. 2008 Sep;40(7):483-93. doi: 10.1002/lsm.20656. Lasers Surg Med. 2008. PMID: 18727026
-
Influence of spatial and temporal spot distribution on the ocular surface quality and maximum ablation depth after photoablation with a 1050 Hz excimer laser system.J Cataract Refract Surg. 2009 Feb;35(2):363-73. doi: 10.1016/j.jcrs.2008.10.053. J Cataract Refract Surg. 2009. PMID: 19185256
-
[Cut quality of a new femtosecond laser system].Klin Monbl Augenheilkd. 2009 Jun;226(6):470-4. doi: 10.1055/s-0028-1109317. Epub 2009 Apr 27. Klin Monbl Augenheilkd. 2009. PMID: 19399716 German.
-
Histological analysis of a cornea following experimental femtosecond laser ablation.Cornea. 2014 Nov;33 Suppl 11:S19-24. doi: 10.1097/ICO.0000000000000251. Cornea. 2014. PMID: 25289720 Review.
-
The ideal excimer beam for refractive surgery.J Refract Surg. 2006 Nov;22(9):S969-72. doi: 10.3928/1081-597X-20061101-27. J Refract Surg. 2006. PMID: 17124899 Review.
Cited by
-
Centration axis in refractive surgery.Eye Vis (Lond). 2015 Feb 24;2:4. doi: 10.1186/s40662-015-0014-6. eCollection 2015. Eye Vis (Lond). 2015. PMID: 26605360 Free PMC article.
-
Comparison of 6.0 mm versus 6.5 mm Optical Zone on Visual Outcomes after LASIK.J Clin Med. 2021 Aug 24;10(17):3776. doi: 10.3390/jcm10173776. J Clin Med. 2021. PMID: 34501222 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical