Feasibility of circumferential pulmonary vein isolation using a novel endoscopic ablation system
- PMID: 20657031
- DOI: 10.1161/CIRCEP.110.954149
Feasibility of circumferential pulmonary vein isolation using a novel endoscopic ablation system
Abstract
Background: Pulmonary vein isolation (PVI) is an established treatment option for patients with drug refractory paroxysmal atrial fibrillation. A novel compliant endoscopic ablation system housing a 980-nm-diode laser allows for discrete point-by-point ablation enabling a true circumferential ablation line design. We sought to determine the feasibility and safety of a circumferential ablation using endoscopic ablation system.
Methods and results: Thirty patients (17 female; mean age, 58±9 years) with a median paroxysmal atrial fibrillation history of 3 years (range, 1 to 17 years) were treated. PVI was achieved in 114 of 116 (98%) PVs (4 left common PVs), thereby achieving simultaneous PVI for separate ipsilateral PVs in 19 of 26 (73%) left PVs and 6 of 30 (20%) right PVs. The total procedure time was 250±62 minutes. Procedure time decreased from 310±59 to 220±37 minutes (P=0.0001) between the first 10 and the last 20 cases. Mean fluoroscopy time was 30±18 minutes. Twenty-seven patients underwent postoperative endoscopy showing no or minimal thermal lesions in the esophagus in 21 (78%) and 2 (7%), respectively. In 4 (15%) patients, an esophageal ulceration was found that healed without sequelae. One pericardial tamponade and 1 right-sided phrenic nerve palsy occurred. During a median follow-up of 168 days (113 to 203 days; q1-q3), 24 of 30 patients (80%) remained free of atrial fibrillation recurrence.
Conclusions: Circumferential PVI using the novel compliant endoscopic ablation system was feasible in the majority of left PVs and minority of right PVs accompanied by a complication rate comparable to established approaches. To minimize the risk for thermal esophageal injury temperature monitoring is recommended.
Similar articles
-
One-year clinical outcome after pulmonary vein isolation using the novel endoscopic ablation system in patients with paroxysmal atrial fibrillation.Heart Rhythm. 2011 Jul;8(7):988-93. doi: 10.1016/j.hrthm.2011.02.030. Epub 2011 Feb 23. Heart Rhythm. 2011. PMID: 21354329
-
Visually guided sequential pulmonary vein isolation: insights into techniques and predictors of acute success.J Cardiovasc Electrophysiol. 2012 Jun;23(6):576-82. doi: 10.1111/j.1540-8167.2011.02247.x. Epub 2012 Jan 9. J Cardiovasc Electrophysiol. 2012. PMID: 22229948
-
Impact of Complete Versus Incomplete Circumferential Lines Around the Pulmonary Veins During Catheter Ablation of Paroxysmal Atrial Fibrillation: Results From the Gap-Atrial Fibrillation-German Atrial Fibrillation Competence Network 1 Trial.Circ Arrhythm Electrophysiol. 2016 Jan;9(1):e003337. doi: 10.1161/CIRCEP.115.003337. Circ Arrhythm Electrophysiol. 2016. PMID: 26763226 Clinical Trial.
-
Endoscopic ablation systems.Expert Rev Med Devices. 2013 Mar;10(2):177-83. doi: 10.1586/erd.12.86. Expert Rev Med Devices. 2013. PMID: 23480087 Review.
-
Minimally invasive surgery for atrial fibrillation.Trends Cardiovasc Med. 2016 Apr;26(3):268-77. doi: 10.1016/j.tcm.2015.07.004. Epub 2015 Jul 20. Trends Cardiovasc Med. 2016. PMID: 26296538 Review.
Cited by
-
New Technologies In Atrial Fibrillation Ablation.J Atr Fibrillation. 2014 Aug 31;7(2):1022. doi: 10.4022/jafib.1022. eCollection 2014 Aug-Sep. J Atr Fibrillation. 2014. PMID: 27957091 Free PMC article. Review.
-
Witnessed pulmonary vein arrest-endoscopically observed conversion of atrial fibrillation into sinus rhythm by laser balloon ablation.Herz. 2012 May;37(3):336-7. doi: 10.1007/s00059-011-3520-x. Epub 2011 Nov 11. Herz. 2012. PMID: 22071678
-
Calorimetric analysis of ice onset temperature during cryoablation: a model approach to identify early predictors of effective applications.Sci Rep. 2021 Aug 4;11(1):15798. doi: 10.1038/s41598-021-95204-2. Sci Rep. 2021. PMID: 34349185 Free PMC article.
-
2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation.Europace. 2018 Jan 1;20(1):e1-e160. doi: 10.1093/europace/eux274. Europace. 2018. PMID: 29016840 Free PMC article. Review. No abstract available.
-
Clinical Applications of Laser Technology: Laser Balloon Ablation in the Management of Atrial Fibrillation.Micromachines (Basel). 2021 Feb 12;12(2):188. doi: 10.3390/mi12020188. Micromachines (Basel). 2021. PMID: 33673330 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Research Materials