Liquid chromatography on a monolithic column microfluidic chip coupled with "three-T" sample injection mode and amperometric detection
- PMID: 20658493
- DOI: 10.1002/jssc.201000304
Liquid chromatography on a monolithic column microfluidic chip coupled with "three-T" sample injection mode and amperometric detection
Abstract
An LC microfluidic chip (LC chip) with amperometric detection was developed. The LC chip employed a 7.5 cm long reversed-phase polymethacrylate monolithic column as a stationary phase and a "three-T" injection mode. A convenient interface was designed to conduct pump pressure into the microfluidic chip to drive solution, and a home-made device was used to control the distance between the working electrode and the LC chip accurately. The "three-T" sample injection mode completely avoided the problem of sample dilution and sample leakage during separation, which is usually observed in traditional and "T" type LC chip, without the using of valve and finally results in a better resolution, reproducibility and relatively high sensitivity. Using the proposed LC chip system, we have successfully separated two isomers, catechol and hydroquinone, within 12 min with a RSD (n=3) <3.0% for retention time and <2.4% for peak area. We have also successfully separated and determined 5-hydroxy-L-tryptophan, dopamine and 5-hydroxytryptamine within 25 min with a RSD (n=3) <5% (for peak area) and a detection limit of 0.16-0.51 μmol/L.
Similar articles
-
Valveless gated injection for microfluidic chip-based liquid chromatography system with polymer monolithic column.J Chromatogr A. 2012 Jul 13;1246:123-8. doi: 10.1016/j.chroma.2012.03.045. Epub 2012 Mar 22. J Chromatogr A. 2012. PMID: 22483268
-
Miniaturised medium pressure capillary liquid chromatography system with flexible open platform design using off-the-shelf microfluidic components.Anal Chim Acta. 2015 Oct 8;896:166-76. doi: 10.1016/j.aca.2015.09.015. Epub 2015 Sep 25. Anal Chim Acta. 2015. PMID: 26482001
-
Reversed-phase liquid chromatography on a microchip with sample injector and monolithic silica column.J Chromatogr A. 2006 Nov 3;1132(1-2):90-8. doi: 10.1016/j.chroma.2006.07.025. Epub 2006 Jul 28. J Chromatogr A. 2006. PMID: 16876806
-
Recent developments in microfluidic chip-based separation devices coupled to MS for bioanalysis.Bioanalysis. 2013 Oct;5(20):2567-80. doi: 10.4155/bio.13.196. Bioanalysis. 2013. PMID: 24138628 Review.
-
Implementations of two-dimensional liquid chromatography.J Chromatogr A. 2008 May 2;1189(1-2):109-68. doi: 10.1016/j.chroma.2008.01.086. Epub 2008 Feb 16. J Chromatogr A. 2008. PMID: 18336826 Review.
Cited by
-
Advances in monoliths and related porous materials for microfluidics.Biomicrofluidics. 2016 May 4;10(3):032901. doi: 10.1063/1.4948507. eCollection 2016 May. Biomicrofluidics. 2016. PMID: 27190564 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources