Postprandial glycemic excursions with the use of a closed-loop platform in subjects with type 1 diabetes: a pilot study
- PMID: 20663458
- PMCID: PMC2909526
- DOI: 10.1177/193229681000400423
Postprandial glycemic excursions with the use of a closed-loop platform in subjects with type 1 diabetes: a pilot study
Abstract
Background: The aim of this study was to evaluate the efficacy of a proportional derivative algorithm closed-loop system to control postprandial glucose concentrations in subjects with type 1 diabetes.
Methods: Six subjects treated with continuous subcutaneous insulin infusion received a standardized meal on three days. The first day served as control, the second day as learning experiment for the algorithm, and the third day to compare the closed loop to the control day. Venous blood glucose was measured as reference until 300 min postprandially. The artificial pancreas platform consisted of a subcutaneous continuous glucose monitor (CGM), the GlucoDay S (Menarini Diagnostics), two D-Tron+ pumps (Disetronic Medical Systems) for subcutaneous insulin, and glucagon administration connected to a personal computer.
Results: One subject was excluded due to technical failure of the CGM. Two of five subjects were male, mean age was 50.8 years (range 38-60), and mean hemoglobin A1c was 8.7% (range 7.0-12.2). The mean postprandial venous blood glucose concentration of day 1 was 205 mg/dl (range 94-265 mg/dl) compared with 128 mg/dl (range 128-158 mg/dl) on day 3 (p = .14). Percentage of time spent in euglycemia postprandially on day 1 was 31% versus 60% on day 3 (p = .08). Time spent below 3.9 mmol/liter (70 mg/dl) was 19% on day 1 compared with 11% on day 3 (p = 1.0). Time above 10 mmol/liter (180 mg/dl) on day 1 was 60% versus 29% on day 3 (p = .22).
Conclusion: The artificial pancreas provided comparable postprandial glycemic control to usual care.
2010 Diabetes Technology Society.
Figures
Similar articles
-
Closed-loop artificial pancreas using subcutaneous glucose sensing and insulin delivery and a model predictive control algorithm: the Virginia experience.J Diabetes Sci Technol. 2009 Sep 1;3(5):1031-8. doi: 10.1177/193229680900300506. J Diabetes Sci Technol. 2009. PMID: 20144416 Free PMC article.
-
Feasibility of a bihormonal closed-loop system to control postexercise and postprandial glucose excursions.J Diabetes Sci Technol. 2012 Sep 1;6(5):1114-22. doi: 10.1177/193229681200600516. J Diabetes Sci Technol. 2012. PMID: 23063038 Free PMC article. Clinical Trial.
-
Closed-loop artificial pancreas using subcutaneous glucose sensing and insulin delivery and a model predictive control algorithm: preliminary studies in Padova and Montpellier.J Diabetes Sci Technol. 2009 Sep 1;3(5):1014-21. doi: 10.1177/193229680900300504. J Diabetes Sci Technol. 2009. PMID: 20144414 Free PMC article.
-
Current topics in glycemic control by wearable artificial pancreas or bedside artificial pancreas with closed-loop system.J Artif Organs. 2016 Sep;19(3):209-18. doi: 10.1007/s10047-016-0904-y. Epub 2016 May 3. J Artif Organs. 2016. PMID: 27142278 Review.
-
Safe glycemic management during closed-loop treatment of type 1 diabetes: the role of glucagon, use of multiple sensors, and compensation for stress hyperglycemia.J Diabetes Sci Technol. 2011 Nov 1;5(6):1373-80. doi: 10.1177/193229681100500608. J Diabetes Sci Technol. 2011. PMID: 22226254 Free PMC article. Review.
Cited by
-
Realizing a Closed-Loop (Artificial Pancreas) System for the Treatment of Type 1 Diabetes.Endocr Rev. 2019 Dec 1;40(6):1521-1546. doi: 10.1210/er.2018-00174. Endocr Rev. 2019. PMID: 31276160 Free PMC article. Review.
-
Design and Evaluation of a Robust PID Controller for a Fully Implantable Artificial Pancreas.Ind Eng Chem Res. 2015 Oct 28;54(42):10311-10321. doi: 10.1021/acs.iecr.5b01237. Epub 2015 Jun 9. Ind Eng Chem Res. 2015. PMID: 26538805 Free PMC article.
-
Impact of erroneous meal insulin bolus with dual-hormone artificial pancreas using a simplified bolus strategy - A randomized controlled trial.Sci Rep. 2018 Feb 8;8(1):2621. doi: 10.1038/s41598-018-20785-4. Sci Rep. 2018. PMID: 29422651 Free PMC article. Clinical Trial.
-
Hypoglycemia Detection and Carbohydrate Suggestion in an Artificial Pancreas.J Diabetes Sci Technol. 2016 Nov 1;10(6):1236-1244. doi: 10.1177/1932296816658666. Print 2016 Nov. J Diabetes Sci Technol. 2016. PMID: 27464755 Free PMC article. Clinical Trial.
-
Autonomous and continuous adaptation of a bihormonal bionic pancreas in adults and adolescents with type 1 diabetes.J Clin Endocrinol Metab. 2014 May;99(5):1701-11. doi: 10.1210/jc.2013-4151. Epub 2014 Jan 31. J Clin Endocrinol Metab. 2014. PMID: 24483160 Free PMC article. Clinical Trial.
References
-
- Hovorka R, Kremen J, Blaha J, Matias M, Anderlova K, Bosanska L, Roubicek T, Wilinska ME, Chassin LJ, Svacina S, Haluzik M. Blood glucose control by a model predictive control algorithm with variable sampling rate versus a routine glucose management protocol in cardiac surgery patients: a randomized controlled trial. J Clin Endocrinol Metab. 2007;92(8):2960–2964. - PubMed
-
- Plank J, Blaha J, Cordingley J, Wilinska ME, Chassin LJ, Morgan C, Squire S, Haluzik M, Kremen J, Svacina S, Toller W, Plasnik A, Ellmerer M, Hovorka R, Pieber TR. Multicentric, randomized, controlled trial to evaluate blood glucose control by the model predictive control algorithm versus routine glucose management protocols in intensive care unit patients. Diabetes Care. 2006;29(2):271–276. - PubMed
-
- Steil GM, Rebrin K, Darwin C, Hariri F, Saad MF. Feasibility of automating insulin delivery for the treatment of type 1 diabetes. Diabetes. 2006;55(12):3344–3350. - PubMed
-
- Weinzimer SA, Steil GM, Swan KL, Dziura J, Kurtz N, Tamborlane WV. Fully automated closed-loop insulin delivery versus semiautomated hybrid control in pediatric patients with type 1 diabetes using an artificial pancreas. Diabetes Care. 2008;31(5):934–939. - PubMed
-
- Schaller HC, Schaupp L, Bodenlenz M, Wilinska ME, Chassin LJ, Wach P, Vering T, Hovorka R, Pieber TR. On-line adaptive algorithm with glucose prediction capacity for subcutaneous closed loop control of glucose: evaluation under fasting conditions in patients with diabetes type 1 diabetes. Diabet Med. 2006;23(1):90–93. - PubMed
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical