Determination of the curling behavior of a preformed cochlear implant electrode array
- PMID: 20665247
- DOI: 10.1007/s11548-010-0520-x
Determination of the curling behavior of a preformed cochlear implant electrode array
Abstract
Purpose: Accurate insertion of a cochlear implant electrode array into the cochlea's helical shape is a crucial step for residual hearing preservation. In image-guided surgery, especially using an automated insertion tool, the overall accuracy of the operative procedure can be improved by adapting the electrode array's intracochlear movement to the individual cochlear shape.
Methods: The curling characteristic of a commercially available state-of-the-art preformed electrode array (Cochlear Ltd. Contour Advance(TM) Electrode Array) was determined using an image-processing algorithm to detect its shape in series of images. An automatic image-processing procedure was developed using Matlab and the Image Processing Toolbox (MathWorks, Natick, Massachusetts, USA) to determine the complete curvature of the electrode array by identifying the 22 platinum contacts of the electrode. A logarithmic spiral was used for a comprehensive mathematical description of the shape of the electrode array. A fitting algorithm for nonlinear least-squares problems was used to provide a complete mathematical description of the electrode array. The system was tested for curling behavior as a function of stylet extraction using nine Contour Advance Research Electrodes (RE) and additionally for nine Contour Advance Practice Electrodes (PE).
Results: All arrays show a typical pattern of curling with adequate predictability after the first 2 or 3 millimeters of stylet extraction. Although non-negligible variations in the overall curling behavior were detected, the electrode arrays show a characteristic movement due to the stylet extraction and only vary minimally after this initial phase.
Conclusion: These results indicate that the risk of intracochlear trauma can be reduced if the specific curling behavior of the electrode carrier is incorporated into the insertion algorithm. Furthermore, the determination of the curling behavior is an essential step in computer-aided cochlear implant electrode development. Experimental data are required for accurate evaluation of the simulation model.
Similar articles
-
Automated insertion of preformed cochlear implant electrodes: evaluation of curling behaviour and insertion forces on an artificial cochlear model.Int J Comput Assist Radiol Surg. 2010 Mar;5(2):173-81. doi: 10.1007/s11548-009-0299-9. Epub 2009 Apr 10. Int J Comput Assist Radiol Surg. 2010. PMID: 20033522
-
An automated insertion tool for cochlear implants: another step towards atraumatic cochlear implant surgery.Int J Comput Assist Radiol Surg. 2010 Mar;5(2):163-71. doi: 10.1007/s11548-009-0368-0. Epub 2009 Jun 13. Int J Comput Assist Radiol Surg. 2010. PMID: 20033518
-
An overview of cochlear implant electrode array designs.Hear Res. 2017 Dec;356:93-103. doi: 10.1016/j.heares.2017.10.005. Epub 2017 Oct 18. Hear Res. 2017. PMID: 29102129 Review.
-
Evaluation of the advance off-stylet insertion technique and the cochlear insertion tool in temporal bones.Otol Neurotol. 2005 Nov;26(6):1161-70. doi: 10.1097/01.mao.0000179527.17285.85. Otol Neurotol. 2005. PMID: 16272935
-
Robotics, automation, active electrode arrays, and new devices for cochlear implantation: A contemporary review.Hear Res. 2022 Feb;414:108425. doi: 10.1016/j.heares.2021.108425. Epub 2021 Dec 25. Hear Res. 2022. PMID: 34979455 Review.
Cited by
-
Individual Optimization of the Insertion of a Preformed Cochlear Implant Electrode Array.Int J Otolaryngol. 2015;2015:724703. doi: 10.1155/2015/724703. Epub 2015 Sep 10. Int J Otolaryngol. 2015. PMID: 26448764 Free PMC article.
-
Advances to electrode pullback in cochlear implant surgery.ScientificWorldJournal. 2012;2012:126767. doi: 10.1100/2012/126767. Epub 2012 Oct 22. ScientificWorldJournal. 2012. PMID: 23193376 Free PMC article.
-
Variations in microanatomy of the human modiolus require individualized cochlear implantation.Sci Rep. 2022 Mar 23;12(1):5047. doi: 10.1038/s41598-022-08731-x. Sci Rep. 2022. PMID: 35322066 Free PMC article.
-
Why Pre-Curved Modiolar Hugging Electrodes Only Cover The Basal Turn of The Cochlea and Not Beyond that?J Int Adv Otol. 2018 Dec;14(3):376-381. doi: 10.5152/iao.2018.5831. J Int Adv Otol. 2018. PMID: 30644378 Free PMC article.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources