Percutaneous transluminal dilatation of inadvertent partial or complete occlusion of the femoral artery caused by Angio-Seal deployment for puncture site closure after cardiac catheterization
- PMID: 20679669
Percutaneous transluminal dilatation of inadvertent partial or complete occlusion of the femoral artery caused by Angio-Seal deployment for puncture site closure after cardiac catheterization
Abstract
The use of access-site closure devices in interventional cardiology has expanded rapidly in the past several years. Initial reports indicated remarkable safety with these devices but there are a small number of major complications. This single-center, retrospective study was carried out to investigate the efficacy of percutaneous transluminal angioplasty (PTA) of inadvertent occlusion or severe stenosis of the femoral artery after deployment of the Angio-Seal closure device. From April 1995 to August 2009 a total of 8,587 Angio-Seal devices were deployed immediately after cardiac catheterization. Within 7 days after deployment, clinical signs of acute arterial occlusion or severe stenosis of the femoral artery were evident in 6 patients (0.07%). They were immediately referred for diagnostic angiography and subsequent PTA. Angiography revealed occlusion of the femoral artery at the puncture site level in 4 patients and a severe stenosis in 2 patients. Using a cross-over access from the opposite groin, PTA of the closed or highly stenosed femoral artery was feasible in all cases. At routine follow up 12 months thereafter, there was no clinical evidence of symptomatic restenosis. Inadvertent occlusion or high-grade stenosis of the femoral artery after Angio-Seal closure device deployment is a rare but severe complication. PTA appears to be a safe and effective method of treatment, avoiding surgical removal of the plug.
Comment in
-
Unintended consequences of femoral artery closure devices.J Invasive Cardiol. 2010 Aug;22(8):358-9. J Invasive Cardiol. 2010. PMID: 20679670 No abstract available.
Similar articles
-
Treatment of Angio-Seal-Related Femoral Artery Occlusion Using Directional Atherectomy-Primary Results and Midterm Follow-Up.Vasc Endovascular Surg. 2016 Nov;50(8):541-546. doi: 10.1177/1538574416679520. Epub 2016 Nov 16. Vasc Endovascular Surg. 2016. PMID: 27852881
-
Excimer laser-assisted recanalisation of femoral arterial stenosis or occlusion caused by the use of Angio-Seal.Eur Radiol. 2001;11(8):1364-70. doi: 10.1007/s003300000793. Eur Radiol. 2001. PMID: 11519544
-
Complications associated with the arterial puncture closure device--Angio-Seal.Vasc Endovascular Surg. 2008 Jun-Jul;42(3):225-7. doi: 10.1177/1538574407312657. Epub 2008 Jan 29. Vasc Endovascular Surg. 2008. PMID: 18230871
-
Vascular occlusion as a late complication of the Angio-Seal closure device. A review of literature.Minerva Cardioangiol. 2007 Dec;55(6):815-9. Minerva Cardioangiol. 2007. PMID: 18091648 Review.
-
Arterial thrombosis after using Angio-Seal.Ann Vasc Surg. 2011 Nov;25(8):1078-93. doi: 10.1016/j.avsg.2010.11.021. Ann Vasc Surg. 2011. PMID: 22023942 Review.
Cited by
-
A retrospective review on feasibility and safety of a new pneumatic compression device for femoral arteriotomy hemostasis.Korean J Radiol. 2012 Jan-Feb;13(1):61-5. doi: 10.3348/kjr.2012.13.1.61. Epub 2011 Dec 23. Korean J Radiol. 2012. PMID: 22247637 Free PMC article.
-
Comparison of Exo-Seal(®) and Angio-Seal (®) for arterial puncture site closure: A randomized, multicenter, single-blind trial.Herz. 2015 Aug;40(5):809-16. doi: 10.1007/s00059-015-4306-3. Epub 2015 Jun 13. Herz. 2015. PMID: 26070467 Clinical Trial.
-
Usefulness and safety of the "God's Hand" pneumatic compression device for hemostasis in femoral catheterization.Diagn Interv Radiol. 2017 Jan-Feb;23(1):49-54. doi: 10.5152/dir.2016.15451. Diagn Interv Radiol. 2017. PMID: 27856404 Free PMC article.