Effect of tracheal tube cuff shape on fluid leakage across the cuff: an in vitro study
- PMID: 20682571
- DOI: 10.1093/bja/aeq202
Effect of tracheal tube cuff shape on fluid leakage across the cuff: an in vitro study
Abstract
Background: This study compared the fluid leakage in the new 'tapered' shaped against the classic 'cylindrical' shaped tracheal tube cuffs when placed in different sized tracheas.
Methods: The 7.5 mm internal diameter (ID) tracheal tube cuffs-Tapered Seal Guard (TSG), Standard Seal Guard (SSG), Hi-Lo, Microcuff, Ruesch, and Portex Profile-were compared in an in vitro apparatus. Vertical artificial tracheas with 16, 20, and 22 mm ID were intubated, 5 ml clear water was applied above the unlubricated tube cuffs, and fluid leakage was measured up to 60 min. Data of tapered vs non-tapered tube cuffs (16 observations) were compared for each tracheal diameter using the Mann-Whitney test.
Results: Median (range) fluid leakage (ml) at 60 min was 2.14 (0.05-4.88), 1.14 (0.00-4.84), and 0.13 (0.00-1.32), respectively, for 16, 20, and 22 mm tracheas in the TSG tube studies when compared with 4.58 (0.44-4.88), 2.21 (0.00-4.81), and 0.00 (0.00-4.81) in the SSG tube and 4.54 (1.54-4.82), 0.90 (0.00-4.49), and 4.85 (4.40-4.99) in the Microcuff tube studies. Leakage in all polyvinylchloride (PVC) tube cuffs was almost complete (5 ml) within 5 min (P<0.001).
Conclusions: The tapered PU tube cuff was as effective as the cylindrical PU cuffs in smaller tracheal diameters and was more efficient than the cylindrical Microcuff PU tube cuff in larger tracheal diameter in preventing subglottic fluid leakage across the tube cuff tested in this in vitro study. PVC tube cuffs leaked much more and faster than PU cuffs.
Similar articles
-
Comparison of air-sealing characteristics of tapered- vs. cylindrical-shaped high-volume, low-pressure tube cuffs.Acta Anaesthesiol Scand. 2012 Feb;56(2):230-5. doi: 10.1111/j.1399-6576.2011.02542.x. Epub 2011 Oct 14. Acta Anaesthesiol Scand. 2012. PMID: 22091784
-
Fluid leakage past tracheal tube cuffs: evaluation of the new Microcuff endotracheal tube.Intensive Care Med. 2003 Oct;29(10):1849-53. doi: 10.1007/s00134-003-1933-6. Epub 2003 Aug 16. Intensive Care Med. 2003. PMID: 12923620
-
Effect of Lanz pressure regulating valve on self-sealing mechanism and air leakage across the tracheal tube cuffs in a benchtop model.J Intensive Care Med. 2013 Jul-Aug;28(4):247-51. doi: 10.1177/0885066612452847. Epub 2012 Jul 24. J Intensive Care Med. 2013. PMID: 22833049
-
Comparison of polyurethane tracheal tube cuffs and conventional polyvinyl chloride tube cuff for prevention of ventilator-associated pneumonia: A systematic review with meta-analysis.Medicine (Baltimore). 2021 Mar 5;100(9):e24906. doi: 10.1097/MD.0000000000024906. Medicine (Baltimore). 2021. PMID: 33655952 Free PMC article.
-
Prevention of Ventilator-Associated and Early Postoperative Pneumonia Through Tapered Endotracheal Tube Cuffs: A Systematic Review and Meta-Analysis of Randomized Controlled Trials.Crit Care Med. 2018 Feb;46(2):316-323. doi: 10.1097/CCM.0000000000002889. Crit Care Med. 2018. PMID: 29206767
Cited by
-
Devising negative pressure within intercuff space reduces microaspiration.BMC Anesthesiol. 2018 Dec 3;18(1):181. doi: 10.1186/s12871-018-0643-0. BMC Anesthesiol. 2018. PMID: 30509183 Free PMC article.
-
Optimal care and design of the tracheal cuff in the critically ill patient.Ann Intensive Care. 2014 Feb 27;4(1):7. doi: 10.1186/2110-5820-4-7. Ann Intensive Care. 2014. PMID: 24572178 Free PMC article.
-
Novel preventive strategies for ventilator-associated pneumonia.Crit Care. 2012 Dec 12;16(2):210. doi: 10.1186/cc11225. Crit Care. 2012. PMID: 22429668 Free PMC article. Review. No abstract available.
-
Comparison of the cuff pressures of a TaperGuard endotracheal tube during ipsilateral and contralateral rotation of the head: A randomized prospective study.Medicine (Baltimore). 2018 Oct;97(42):e12702. doi: 10.1097/MD.0000000000012702. Medicine (Baltimore). 2018. PMID: 30334954 Free PMC article. Clinical Trial.
-
Comparison of the cuff pressure of a TaperGuard endotracheal tube and a cylindrical endotracheal tube after lateral rotation of head during middle ear surgery: A single-blind, randomized clinical consort study.Medicine (Baltimore). 2017 Mar;96(10):e6257. doi: 10.1097/MD.0000000000006257. Medicine (Baltimore). 2017. PMID: 28272230 Free PMC article. Clinical Trial.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous