. 2010 Sep;36(9):1517-22.
doi: 10.1016/j.jcrs.2010.02.028.

Intraocular stability of an angle-supported phakic intraocular lens with changes in pupil diameter

Affiliations

Intraocular stability of an angle-supported phakic intraocular lens with changes in pupil diameter

Jorge L Alió et al. J Cataract Refract Surg. 2010 Sep.

Abstract

Purpose: To use anterior segment optical coherence tomography (AS-OCT) to evaluate the stability of a recently released angle-supported phakic intraocular lens (pIOL) in the anterior segment with changes in pupil diameter.

Setting: Keratoconus Unit, Vissum Corporation, Alicante, Spain.

Methods: In this observational cross-sectional study of consecutive eyes with moderate to high myopia, an AcrySof Cachet pIOL was implanted with the aim of minimizing the refractive error. An analysis of the position and stability of the pIOL before and after pharmacologic pupil dilation was performed 3 months postoperatively using the Visante AS-OCT system. A measurement protocol that included several anatomic parameters was developed and applied; the parameter values before and after dilation were compared.

Results: Twenty eyes of 20 patients ranging in age from 24 to 48 years old were evaluated. The anterior chamber depth increased significantly with pupil dilation (mean change 0.06 mm +/- 0.08 [SD]) (P<.01). A significant change was also observed in the distance between the center of the cornea at the endothelial plane and the anterior surface of the pIOL (mean change 0.03 +/- 0.05 mm) (P = .01). The distances between the peripheral edges of the pIOL and the corneal endothelium and the distance between the crystalline lens and the pIOL did not change significantly (P>or=.14).

Conclusion: The angle-supported pIOL showed excellent intraocular behavior after pupil dilation, with no shortening of the distance between the pIOL and corneal endothelium at the center or peripheral edges of the pIOL.