Calibration function for the Orbitrap FTMS accounting for the space charge effect

Affiliations
Free article

Calibration function for the Orbitrap FTMS accounting for the space charge effect

Mikhail V Gorshkov et al. J Am Soc Mass Spectrom. 2010 Nov.
Free article

Abstract

Ion storage in an electrostatic trap has been implemented with the introduction of the Orbitrap Fourier transform mass spectrometer (FTMS), which demonstrates performance similar to high-field ion cyclotron resonance MS. High mass spectral characteristics resulted in rapid acceptance of the Orbitrap FTMS for Life Sciences applications. The basics of Orbitrap operation are well documented; however, like in any ion trap MS technology, its performance is limited by interactions between the ion clouds. These interactions result in ion cloud couplings, systematic errors in measured masses, interference between ion clouds of different size yet with close m/z ratios, etc. In this work, we have characterized the space-charge effect on the measured frequency for the Orbitrap FTMS, looking for the possibility to achieve sub-ppm levels of mass measurement accuracy (MMA) for peptides in a wide range of total ion population. As a result of this characterization, we proposed an m/z calibration law for the Orbitrap FTMS that accounts for the total ion population present in the trap during a data acquisition event. Using this law, we were able to achieve a zero-space charge MMA limit of 80 ppb for the commercial Orbitrap FTMS system and sub-ppm level of MMA over a wide range of total ion populations with the automatic gain control values varying from 10 to 10(7).

PubMed Disclaimer

Similar articles

Cited by

References

    1. Proc Natl Acad Sci U S A. 1996 Dec 10;93(25):14440-5 - PubMed
    1. Anal Chem. 1984 Dec;56(14):2744-8 - PubMed
    1. J Proteome Res. 2004 Jan-Feb;3(1):61-7 - PubMed
    1. J Am Soc Mass Spectrom. 2004 May;15(5):703-14 - PubMed
    1. Mol Cell Proteomics. 2007 Mar;6(3):377-81 - PubMed

Publication types