Laser Doppler flowmetry: measurements in a layered perfusion model and Monte Carlo simulations of measurements
- PMID: 20700248
- DOI: 10.1364/AO.30.002592
Laser Doppler flowmetry: measurements in a layered perfusion model and Monte Carlo simulations of measurements
Abstract
Fluid flow in a flow model was measured with a laser Doppler flowmeter normally used for tissue perfusion measurements. The flow parameter is deduced from the first moment of the spectral power density of intensity fluctuations at a detector viewing the model, which is illuminated by a laser. The model consisted of layers, filled with liquid, containing scattering particles, at different distances to the optics of the flowmeter. To each of the layers a flow can be applied. The configuration of the optical components could be varied, which resulted in an adjustable relative sensitivity for the flow in the different layers. Here, the feasibility of depth-dependent flowmetry with the laser Doppler flowmeter is shown experimentally. The measurements in the flow model were simulated with Monte Carlo calculations. The results indicate that these calculations may be of help to interpret experimentally obtained flow results.
Similar articles
-
Monte Carlo simulations of laser Doppler blood flow measurements in tissue.Appl Opt. 1990 Jun 1;29(16):2371-81. doi: 10.1364/AO.29.002371. Appl Opt. 1990. PMID: 20563178
-
Doppler Monte Carlo simulations of light scattering in tissue to support laser-Doppler perfusion measurements.Technol Health Care. 1999;7(2-3):171-83. Technol Health Care. 1999. PMID: 10463306
-
Laser Doppler blood flowmetry using two wavelengths: Monte Carlo simulations and measurements.Appl Opt. 1994 Jun 1;33(16):3549-58. doi: 10.1364/AO.33.003549. Appl Opt. 1994. PMID: 20885742
-
Evaluation of algorithms for microperfusion assessment by fast simulations of laser Doppler power spectral density.Phys Med Biol. 2011 Dec 21;56(24):7709-23. doi: 10.1088/0031-9155/56/24/002. Epub 2011 Nov 16. Phys Med Biol. 2011. PMID: 22085805
-
[Laser Doppler blood flowmetry].Tidsskr Nor Laegeforen. 1991 Oct 10;111(24):2966-8. Tidsskr Nor Laegeforen. 1991. PMID: 1948903 Review. Norwegian.
Cited by
-
Laser speckle contrast imaging of the skin: interest in processing the perfusion data.Med Biol Eng Comput. 2012 Feb;50(2):103-5. doi: 10.1007/s11517-011-0856-6. Epub 2011 Dec 29. Med Biol Eng Comput. 2012. PMID: 22205575
-
Multichannel laser-Doppler probe for blood perfusion measurements with depth discrimination.Med Biol Eng Comput. 1998 Nov;36(6):740-7. doi: 10.1007/BF02518878. Med Biol Eng Comput. 1998. PMID: 10367466
-
Voluntary locomotion linked with cerebral activation is mediated by vasopressin V1a receptors in free-moving mice.J Physiol. 2013 Jul 15;591(14):3651-65. doi: 10.1113/jphysiol.2013.251876. Epub 2013 May 13. J Physiol. 2013. PMID: 23671158 Free PMC article.