An oscillating sweeping gap test for VMAT quality assurance
- PMID: 20702923
- DOI: 10.1088/0031-9155/55/17/010
An oscillating sweeping gap test for VMAT quality assurance
Abstract
The objective of this study was to develop an oscillating sweeping gap test for volumetric modulated arc therapy (VMAT) quality assurance (QA). A novel test was designed and used to simultaneously determine uncertainties associated with linac performance, dose calculation and dosimetric MLC parameters during VMAT delivery. Delivered doses were measured with Matrixx, ionization chamber A12 and EDR2 films, and compared to calculations from the treatment planning system (TPS) Eclipse. A new gantry and MLC motion pattern, called here 'oscillating sweeping gap', is developed as an extension of the standard sweeping gap MLC pattern developed for IMRT QA. Specifically, in the oscillating sweeping gap test, a uniform MLC gap is moving repeatedly back and forth across the field at a constant speed during a full rotation of the gantry. The dose distribution generated by the combined gantry and MLC motion pattern is designed to be quasi-uniform within a cylindrical target volume with a sharp penumbra. The test design allows for an easy detection of dose errors as deviations from the uniform background. MLC gap sizes, gantry and MLC speeds and monitor units (MU) are selected according to a formula determining the magnitude of dose delivered to the target. Both measured and calculated dose distributions were analyzed as a function of the number of control points in the TPS, MLC gap size and magnitude of the gantry angle error. Dose calculation errors due to the insufficient number of control points in the gantry and MLC motion pattern appear as streak artifacts. The magnitude of these artifacts is increasing with the decreased number of control points, and with the decreased MLC gap size. The spatial distribution of dose errors due to the gantry angle errors (unsteady rocking motion) appears as high-frequency noise for higher wobble frequencies and as large hot/cold spots for lower wobble frequencies. The actual MLC leaf position as a function of time (or the gantry angle), determined from the Matrixx snaps (dose images measured per time interval) of the moving gap and compared to the ideal leaf positions, reveal discrepancies in agreement with theoretical calculations. The MLC parameters determined for VMAT with the oscillating sweeping gap test, their uncertainties and the associated dose errors are similar to those determined for IMRT with the standard sweeping gap test. The oscillating sweeping gap test has been developed for the gantry and MLC QA. Applications include commissioning of the planning system for VMAT and performing routine linac QA. The test is sensitive to several errors in dose calculation and delivery.
Similar articles
-
A novel method for routine quality assurance of volumetric-modulated arc therapy.Med Phys. 2013 Oct;40(10):101712. doi: 10.1118/1.4820439. Med Phys. 2013. PMID: 24089902
-
A real time dose monitoring and dose reconstruction tool for patient specific VMAT QA and delivery.Med Phys. 2012 Dec;39(12):7194-204. doi: 10.1118/1.4764482. Med Phys. 2012. PMID: 23231270
-
Quality assurance of volumetric modulated arc therapy using Elekta Synergy.Acta Oncol. 2009;48(8):1193-7. doi: 10.3109/02841860903081905. Acta Oncol. 2009. PMID: 19863228
-
Dosimetric validation for multileaf collimator-based intensity-modulated radiotherapy: a review.Med Dosim. 2001 Summer;26(2):179-88. doi: 10.1016/s0958-3947(01)00058-9. Med Dosim. 2001. PMID: 11444520 Review.
-
Radiochromic film dosimetry: past, present, and future.Phys Med. 2011 Jul;27(3):122-34. doi: 10.1016/j.ejmp.2010.10.001. Epub 2010 Nov 2. Phys Med. 2011. PMID: 21050785 Review.
Cited by
-
Evaluation of QA software system analysis for the static picket fence test.J Appl Clin Med Phys. 2022 Jul;23(7):e13618. doi: 10.1002/acm2.13618. Epub 2022 May 15. J Appl Clin Med Phys. 2022. PMID: 35570379 Free PMC article.
-
Quality control of VMAT synchronization using portal imaging.J Appl Clin Med Phys. 2015 Jan 8;16(1):5238. doi: 10.1120/jacmp.v16i1.5238. J Appl Clin Med Phys. 2015. PMID: 25679179 Free PMC article.
-
On the need for tuning the dosimetric leaf gap for stereotactic treatment plans in the Eclipse treatment planning system.J Appl Clin Med Phys. 2019 Jul;20(7):68-77. doi: 10.1002/acm2.12656. Epub 2019 Jun 21. J Appl Clin Med Phys. 2019. PMID: 31225938 Free PMC article.
-
On using the dosimetric leaf gap to model the rounded leaf ends in VMAT/RapidArc plans.J Appl Clin Med Phys. 2014 Mar 6;15(2):4484. doi: 10.1120/jacmp.v15i2.4484. J Appl Clin Med Phys. 2014. PMID: 24710433 Free PMC article.
-
A novel and independent method for time-resolved gantry angle quality assurance for VMAT.J Appl Clin Med Phys. 2017 Sep;18(5):134-142. doi: 10.1002/acm2.12129. Epub 2017 Jul 13. J Appl Clin Med Phys. 2017. PMID: 28703451 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous