Implementation of a multi-functional ambulatory urodynamics monitoring system based on newly devised abdominal pressure measurement
- PMID: 20703606
- DOI: 10.1007/s10916-009-9318-1
Implementation of a multi-functional ambulatory urodynamics monitoring system based on newly devised abdominal pressure measurement
Abstract
The measurement of the rectal pressure is considered to be the 'gold standard' for the assessment of the abdominal pressure. However, conventional rectal catheters can cause erroneous results and are not comfortable for the patients. To reduce these problems, we devised a non-invasive technique for the measurement of the abdominal pressure using the parametric curve fitting method, based on linear, polynomial, exponential or sine equation modeling, between the rectal pressure and electromyographic (sEMG) signals recorded simultaneously from the abdomen. The sEMG signals and rectal pressure were obtained simultaneously from 12 patients with neurogenic bladders due to spinal cord injury (age = 53.2 ± 11.9 years, BMI = 24.4 ± 2.7, ASIA classification: D). Using our algorithm, the correlation coefficient and root mean square error (RMSE) between the measured and estimated abdominal pressure obtained by the quartic polynomial modeling were 0.86 ± 0.05 and 4.70 ± 1.56, respectively. The results obtained herein suggest that the sEMG signals can be used reliably for the indirect measurement of the abdominal pressure in ambulatory urodynamics monitoring systems.
Similar articles
-
Availability of a newly devised ambulatory urodynamics monitoring system based on personal device assistance in patients with spinal cord injury.Comput Methods Programs Biomed. 2012 Jun;106(3):260-73. doi: 10.1016/j.cmpb.2010.10.008. Epub 2010 Nov 10. Comput Methods Programs Biomed. 2012. PMID: 21071108
-
Intra-abdominal pressure measurement method via the urinary-tube: bedside validation of a biomechanical model integrating urine column height and bladder urinary volume.Ann Surg. 2010 Jan;251(1):127-32. doi: 10.1097/SLA.0b013e3181b5919b. Ann Surg. 2010. PMID: 19864939
-
Intra-abdominal pressure measurement using a U-tube technique: caveat emptor!Ann Surg. 2010 Nov;252(5):890; author reply 890-1. doi: 10.1097/SLA.0b013e3181fded35. Ann Surg. 2010. PMID: 21037448 No abstract available.
-
[Significance of rectal contractions during multichannel urodynamic study].Nihon Hinyokika Gakkai Zasshi. 1997 Oct;88(10):874-9. doi: 10.5980/jpnjurol1989.88.874. Nihon Hinyokika Gakkai Zasshi. 1997. PMID: 9388367 Review. Japanese.
-
Is the canary still singing?Crit Care Med. 2007 Jan;35(1):320-1. doi: 10.1097/01.CCM.0000251843.34578.F2. Crit Care Med. 2007. PMID: 17197785 Review. No abstract available.