. 2010 Sep-Oct;15(5):257-61.
doi: 10.1111/j.1478-5153.2010.00422.x.

Increasing positive end expiratory pressure at extubation reduces subglottic secretion aspiration in a bench-top model

Affiliations

Increasing positive end expiratory pressure at extubation reduces subglottic secretion aspiration in a bench-top model

Jack Hodd et al. Nurs Crit Care. 2010 Sep-Oct.

Abstract

Aim: To estimate the ability of simulated tracheal suction, adjusting the positive end expiratory pressure (PEEP) settings on the ventilator or compressing a self-inflating bag to minimize aspiration during cuff deflation and extubation in a bench-top model.

Background: During intubation, colonized secretions accumulate in the subglottic space above the endotracheal tube (ETT) cuff. Consequently, during cuff deflation and extubation, there is a risk of aspiration of the secretions. This may result in pneumonitis or pneumonia. There are a number of techniques used during cuff deflation and extubation to prevent secretion aspiration.

Method: A model trachea was intubated and the proximal end of the ETT was attached to a mechanical ventilator. Ten millilitres of water was placed above the inflated cuff and then nine test protocols were implemented in a random order to simulate tracheal suction, adjusting the PEEP settings on the ventilator or compressing a self-inflating bag. The volume of water 'aspirated' by the model was determined by weighing the apparatus pre- and post-extubation. Statistical analysis was performed using regression analysis and heteroscedastic t tests with a Bonferroni correction.

Results: The level of PEEP was negatively correlated with the volume of fluid aspirated [co-efficient -0.24 (99% confidence interval -0.31 to -0.17), R(2) = 0.75]. Significantly less fluid was aspirated when a PEEP of 35 cmH(2)O was applied when compared with competing techniques.

Discussion and conclusions: This study suggests that applying PEEP during cuff deflation and extubation is protective against aspiration. We conclude that unless there is a contraindication, the application of PEEP should be considered when extubating patients.