. 2011 Jul;38(3):148-55.
doi: 10.1016/j.neurad.2010.05.002. Epub 2010 Aug 21.

Overestimation of moderate carotid stenosis assessed by both Doppler US and contrast enhanced 3D-MR angiography in the CARMEDAS study

Affiliations
Free article

Overestimation of moderate carotid stenosis assessed by both Doppler US and contrast enhanced 3D-MR angiography in the CARMEDAS study

M Nonent et al. J Neuroradiol. 2011 Jul.
Free article

Abstract

Purpose: To evaluate the agreement and diagnostic accuracy of Contrast enhanced magnetic resonance angiography (CE-MRA), Doppler ultrasound (DUS) and Digital subtraction angiography (DSA) in the assessment of carotid stenosis.

Methods: DUS, CE-MRA and DSA were performed in 56 patients included in the Carotide-angiographie par résonance magnétique-échographie-doppler-angioscanner (CARMEDAS) multicenter study with a carotid stenosis ≥ 50%. Three readers evaluated stenoses on CE-MRA and DSA (NASCET criteria). Velocities criteria were used for stenosis estimation on DUS.

Results: CE-MRA had a sensitivity and specificity of 96-98% and 66-83% respectively for carotid stenoses ≥ 50% and a sensitivity and specificity of 94% and 76-84% respectively for carotid stenoses ≥ 70%. The interobserver agreement of CE-MRA was excellent, except for moderate stenoses (50-69%). DUS had a sensitivity and specificity of 88 and 75% respectively for carotid stenoses ≥ 50% and a sensitivity and specificity of 83 and 86% respectively for carotid stenoses ≥ 70%. Combined concordant CE-MRA and DUS had a sensitivity and specificity of 100 and 85-90% respectively for carotid stenoses ≥ 50% and a sensitivity and specificity of 96-100% and 80-87% respectively for carotid stenoses ≥ 70%. The positive predictive value of the association CE-MRA and DUS for carotid stenoses ≥ 70% is calculated between 77 and 82% while the negative predictive value is calculated between 97 and 100%. CE-MRA and DUS have concordant findings in 63-72%, and the overestimations cases were recorded only for carotid stenosis ≤ 69%.

Conclusion: Combined DUS-CE-MRA is excellent for evaluation of severe stenosis but remains debatable in moderate stenosis (50-69%) due to the risk of overestimations.