Retrieval analysis and in vitro assessment of strength, durability, and distraction of a modular total hip replacement
- PMID: 20737432
- DOI: 10.1002/jbm.a.32886
Retrieval analysis and in vitro assessment of strength, durability, and distraction of a modular total hip replacement
Abstract
We investigated a commercial Co-Cr-alloy head--Ti6Al4V alloy neck and Ti6Al4V stem modular total hip replacement. We assessed the distraction forces after in vitro cycling in bovine serum, fatigue durability, fretting corrosion damage, and load bearing capacity of new implants using fatigue-corrosion, pull-off, scanning electron microscopy, fatigue and compression investigations. In addition, we studied corrosion, fretting damage, and distraction forces on retrievals. For both retrievals and in vitro test samples, the neck-stem interface required the higher distraction force as compared with the head-neck interface. One of 12 retrievals showed strong fretting corrosion at the neck-stem interface which resulted in a high disassembly force of about 16 kN. For in vitro test samples, the neck-stem pull-off force initially increased during cycling and showed a maximum value of 5.704 kN at ∼100,000 cycles, which is equivalent to gait cycles performed in approximately 36 days. Overall, assembly force, initial component settling, and interface corrosion primarily determine the force required to distract the modular components. One million cycles fatigue failure of the neck can be expected at a maximum compression load of -6.5 kN. No component failure was observed during quasistatic compression; rather the neck deformed plastically and the ultimate compression load-bearing capacity was -13 kN.
© 2010 Wiley Periodicals, Inc. J Biomed Mater Res Part A, 2010.
Similar articles
-
Differences in the fretting corrosion of metal-metal and ceramic-metal modular junctions of total hip replacements.J Orthop Res. 2004 Mar;22(2):250-9. doi: 10.1016/S0736-0266(03)00186-4. J Orthop Res. 2004. PMID: 15013082
-
An analysis of the head-neck taper interface in retrieved hip prostheses.Clin Orthop Relat Res. 1994 Mar;(300):162-7. Clin Orthop Relat Res. 1994. PMID: 8131330
-
Influence of material coupling and assembly condition on the magnitude of micromotion at the stem-neck interface of a modular hip endoprosthesis.J Biomech. 2011 Jun 3;44(9):1747-51. doi: 10.1016/j.jbiomech.2011.04.007. Epub 2011 Apr 30. J Biomech. 2011. PMID: 21531416
-
Modularity of the femoral component in total hip arthroplasty.J Am Acad Orthop Surg. 2012 Apr;20(4):214-22. doi: 10.5435/JAAOS-20-04-214. J Am Acad Orthop Surg. 2012. PMID: 22474091 Review.
-
[The morse taper junction in modular revision hip replacement].Z Orthop Unfall. 2007 Mar-Apr;145(2):157-60. doi: 10.1055/s-2007-965182. Z Orthop Unfall. 2007. PMID: 17492554 Review. German.
Cited by
-
Corrosion and Tribology of Materials Used in a Novel Reverse Hip Replacement.Materials (Basel). 2017 Jul 5;10(7):751. doi: 10.3390/ma10070751. Materials (Basel). 2017. PMID: 28773112 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical