Autonomic dysreflexia as a complication of a fecal management system in a man with tetraplegia
- PMID: 20737801
- PMCID: PMC2920121
- DOI: 10.1080/10790268.2010.11689705
Autonomic dysreflexia as a complication of a fecal management system in a man with tetraplegia
Abstract
Background/objective: To present a case of autonomic dysreflexia caused by the use of a fecal management system in a patient with tetraplegia.
Design: Case report.
Setting: Military rehabilitation center.
Results: A man with tetraplegia had a fecal management system inserted to divert stool away from his sacral pressure ulcer to reduce contamination and infection risk. Two days later, he developed severe autonomic dysreflexia that improved after removal of the system.
Conclusions: Autonomic dysreflexia, a life-threatening complication, has not been reported before as a side effect of a fecal management system. These systems should be used with caution in patients with high-level spinal cord injury.
Figures
References
-
- Karlsson AK. Autonomic dysreflexia. Spinal Cord. 1999;37(6):383–391. - PubMed
-
- Dolinak D, Balraj E. Autonomic dysreflexia and sudden death in people with traumatic spinal cord injury. Am J Forensic Med Pathol. 2007;28(2):95–98. - PubMed
-
- Flexi-Seal® Fecal Management System [package insert] Princeton, NJ: ConvaTec, a Division of ER Squibb & Sons, LLC; 2006.
-
- Padmanabhan A, Stern M, Wishin J, Mangino M, Richey K, DeSane M. Clinical evaluation of a flexible fecal incontinence management system. Am J Crit Care. 2007;16(4):384–393. - PubMed
-
- Page BP, Boyce SA, Deans C, Camilleri-Brennan J. Significant rectal bleeding as a complication of a fecal collecting device: report of a case. Dis Colon Rectum. 2008;51(9):1427–1429. - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources