Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2010 Sep;43(9):808-17.
doi: 10.1111/j.1365-2591.2010.01765.x. Epub 2010 Aug 3.

Effectiveness of four electronic apex locators to determine distance from the apical foramen

Affiliations
Comparative Study

Effectiveness of four electronic apex locators to determine distance from the apical foramen

R Stoll et al. Int Endod J. 2010 Sep.

Abstract

Aim: To evaluate the accuracy of four electronic apex locators (EAL) in the apical region (0-3 mm short of the foramen) and to compare the precision of the readings on the display with the real position of the file in the root canal.

Methodology: Twenty single-rooted extracted teeth with round root canals were used. The canal orifices were preflared, and the length to the major foramen was determined visually under a microscope. Canals were enlarged, so that a size 15 file fitted well inside the canal. Teeth were mounted in acrylic test tubes filled with physiologic saline solution. Electronic length was determined in the region between the major foramen and 3 mm short of it in 0.5 mm steps with the Dentaport ZX, Root ZX mini, Elements Diagnostic Unit and Apex Locator and Raypex 5 using files of size 10 and size 15. The data were analysed using linear regression between true length and EAL reading, Bland-Altman plots and nonparametric tests at a significance level of alpha = 0.05.

Results: The major foramen was detected by all EALs. With a measurement file positioned directly at the major foramen, meter readings were equivalent to a position 0.01-0.38 mm away. For the Dentaport ZX, a better accuracy using the size 15 file for the area 0-1.5 mm short of the apex was found. The differences in measurements between the two files were smaller for the other EALs. In linear regression, a good linearity for Dentaport ZX and Root ZX mini and moderate linearity for Elements Diagnostic Unit and Apex Locator and Raypex 5 were found. The slope of the measurement curve was too low (0.37-0.57) for the Raypex 5 and almost optimal for the Dentaport ZX (1.01-1.05). The Root ZX mini and the Elements Obturation Unit produced lower slope values and especially the Elements Obturation Unit revealed much higher SDs at the different measurement levels.

Conclusion: Amongst the four EALs, the Dentaport ZX and Root ZX mini had the best agreement between true lengths and meter readings. For the Raypex 5, an interpretation of the colour-coded zones as distance to the foramen cannot be recommended and might lead to erroneous interpretations.

PubMed Disclaimer

Similar articles

Cited by

Publication types

MeSH terms

LinkOut - more resources