Internal rotational error of the tibial component is a major cause of pain after total knee replacement
- PMID: 20798441
- DOI: 10.1302/0301-620X.92B9.23516
Internal rotational error of the tibial component is a major cause of pain after total knee replacement
Abstract
This study used CT analysis to determine the rotational alignment of 39 painful and 26 painless fixed-bearing total knee replacements (TKRs) from a cohort of 740 NexGen Legacy posterior-stabilised and cruciate-retaining prostheses implanted between May 1996 and August 2003. The mean rotation of the tibial component was 4.3 degrees of internal rotation (25.4 degrees internal to 13.9 degrees external rotation) in the painful group and 2.2 degrees of external rotation (8.5 degrees internal to 18.2 degrees external rotation) in the painfree group (p = 0.024). In the painful group 17 tibial components were internally rotated more than 9 degrees compared with none in the painfree group (p < 0.001). Additionally, six femoral components in the painful group were internally rotated more than 6 degrees compared with none in the painfree group (p = 0.017). External rotational errors were not found to be associated with pain. Overall, 22 (56.4%) of the painful TKRs had internal rotational errors involving the femoral, the tibial or both components. It is estimated that at least 4.6% of all our TKRs have been implanted with significant internal rotational errors.
Similar articles
-
Component rotational alignment in unexplained painful primary total knee arthroplasty.Knee. 2014 Jan;21(1):272-7. doi: 10.1016/j.knee.2012.09.011. Epub 2012 Nov 7. Knee. 2014. PMID: 23140906
-
Unexplained pain following total knee arthroplasty: Is rotational malalignment the problem?Knee. 2018 Mar;25(2):329-334. doi: 10.1016/j.knee.2018.01.011. Epub 2018 Mar 9. Knee. 2018. PMID: 29526396
-
Effect of femoral and tibial component position on patellar tracking following total knee arthroplasty: 10-year follow-up of Miller-Galante I knees.Am J Knee Surg. 2001 Summer;14(3):152-6. Am J Knee Surg. 2001. PMID: 11491425
-
Mobile-bearing insert translational and rotational kinematics in a PCL-retaining total knee arthroplasty.Orthop Traumatol Surg Res. 2009 Jun;95(4):254-9. doi: 10.1016/j.otsr.2009.03.012. Epub 2009 May 12. Orthop Traumatol Surg Res. 2009. PMID: 19442597
-
MRI evaluation of rotational alignment and synovitis in patients with pain after total knee replacement.J Bone Joint Surg Br. 2012 Sep;94(9):1209-15. doi: 10.1302/0301-620X.94B9.28489. J Bone Joint Surg Br. 2012. PMID: 22933492
Cited by
-
Maximizing tibial coverage is detrimental to proper rotational alignment.Clin Orthop Relat Res. 2014 Jan;472(1):121-5. doi: 10.1007/s11999-013-3047-y. Clin Orthop Relat Res. 2014. PMID: 23653101 Free PMC article.
-
Kinematic alignment in total knee arthroplasty leads to a better restoration of patellar kinematics compared to mechanic alignment.Knee Surg Sports Traumatol Arthrosc. 2019 May;27(5):1529-1534. doi: 10.1007/s00167-018-5284-9. Epub 2018 Nov 12. Knee Surg Sports Traumatol Arthrosc. 2019. PMID: 30421165
-
Short-Term Benefits of Robotic Assisted Total Knee Arthroplasty Over Computer Navigated Total Knee Arthroplasty Are Not Sustained With No Difference in Postoperative Patient-Reported Outcome Measures.Arthroplast Today. 2022 Jan 12;14:210-215.e0. doi: 10.1016/j.artd.2021.11.014. eCollection 2022 Apr. Arthroplast Today. 2022. PMID: 35510065 Free PMC article.
-
Computed tomography confirmation of component rotation in nanosensor-balanced total knee arthroplasty.Arthroplast Today. 2018 Sep 14;5(1):64-67. doi: 10.1016/j.artd.2018.07.008. eCollection 2019 Mar. Arthroplast Today. 2018. PMID: 31020025 Free PMC article.
-
Design improvement in patient-specific instrumentation for total knee arthroplasty improved the accuracy of the tibial prosthetic alignment in the coronal and axial planes.Knee Surg Sports Traumatol Arthrosc. 2020 May;28(5):1560-1567. doi: 10.1007/s00167-019-05571-7. Epub 2019 Jun 25. Knee Surg Sports Traumatol Arthrosc. 2020. PMID: 31240377