Abstract
Objectives: To establish iodine (I) contrast medium (CM) doses iso-attenuating with gadolinium (Gd) CM doses regarded diagnostic in CTA and percutaneous catheter-angiography/vascular interventions (PCA/PVI) in azotemic patients.
Methods: CT Hounsfield units (HU) were measured in 20-mL syringes containing 0.01/0.02,/0.05/0.1 mmol/mL of iodine or gadolinium atoms and placed in phantoms. Relative contrast were measured in 20-mL syringes filled with iohexol at 35/50/70/90/110/140 mg I/mL and 0.5 M gadodiamide using radiofluoroscopy (RF), digital radiography (DX) and x-ray angiography (XA) systems. Clinical doses of Gd-CM at CTA/PCA/PVI were reviewed.
Results: At CT 91-116 and 104-125 mg I/mL in the chest and abdominal phantoms, respectively, were iso-attenuating with 0.5 M Gd at 80-140 kVp. At RF/DX/XA systems 35-90 mg I/mL were iso-attenuating with 0.5 M gadodiamide at 60-115 kVp. Clinically, 60 mL 91-125 mg I/mL (5.5-7.5 gram-iodine) at 80-140 kVp CTA and 60 mL of 35-90 mg I/mL (2.1-5.4 gram-iodine) at 60-115 kVp PCA/PVI would be iso-attenuating with 60 mL 0.5 M Gd-CM (=0.4 mmol Gd/kg in a 75-kg person).
Conclusions: Meticulous examination technique and judicious use of ultra-low I-CM doses iso-attenuating with diagnostic Gd-CM doses in CTA and PCA/PVI may minimise the risk of nephrotoxicity in azotemic patients, while there is no risk of NSF.
Similar articles
-
Ultra-low iodine concentrations iso-attenuating with diagnostic 0.5M gadolinium in endovascular procedures to minimize the risk of contrast nephropathy: A phantom study.Eur J Radiol. 2015 Jun;84(6):1068-74. doi: 10.1016/j.ejrad.2015.03.014. Epub 2015 Mar 23. Eur J Radiol. 2015. PMID: 25842009
-
Iodine contrast media doses equal-attenuating with gadolinium chelates at CT-aortography may have less risk of contrast-induced nephropathy and no risk of nephrogenic systemic fibrosis in azotaemic patients!Eur Radiol. 2008 Sep;18(9):2013-4. doi: 10.1007/s00330-008-0964-3. Epub 2008 Apr 4. Eur Radiol. 2008. PMID: 18389241 No abstract available.
-
Using the K-edge to improve contrast conspicuity and to lower radiation dose with a 16-MDCT: a phantom and human study.J Comput Assist Tomogr. 2006 May-Jun;30(3):391-7. doi: 10.1097/00004728-200605000-00008. J Comput Assist Tomogr. 2006. PMID: 16778612
-
Are gadolinium-based contrast media really safer than iodinated media for digital subtraction angiography in patients with azotemia?Radiology. 2002 May;223(2):311-8; discussion 328-9. doi: 10.1148/radiol.2232010221. Radiology. 2002. PMID: 11997530 Review.
-
Toxicity of contrast media: an update.Neth J Med. 2008 Nov;66(10):416-22. Neth J Med. 2008. PMID: 19011267 Review.
Cited by
-
Are intravenous injections of contrast media really less nephrotoxic than intra-arterial injections?Eur Radiol. 2012 Jun;22(6):1366-71. doi: 10.1007/s00330-011-2371-4. Epub 2012 Feb 4. Eur Radiol. 2012. PMID: 22307815 Review.
-
Analysis of intra-aneurysmal flow for cerebral aneurysms with cerebral angiography.AJNR Am J Neuroradiol. 2012 Oct;33(9):1679-84. doi: 10.3174/ajnr.A3057. Epub 2012 May 10. AJNR Am J Neuroradiol. 2012. PMID: 22576897 Free PMC article.
-
Waiting times between examinations with intravascularly administered contrast media: a review of contrast media pharmacokinetics and updated ESUR Contrast Media Safety Committee guidelines.Eur Radiol. 2024 Apr;34(4):2512-2523. doi: 10.1007/s00330-023-10085-5. Epub 2023 Oct 12. Eur Radiol. 2024. PMID: 37823923 Free PMC article. Review.
References
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical