Hydroxyapatite-coated pins versus titanium alloy pins in external fixation at the wrist: a controlled cohort study
- PMID: 20805762
- DOI: 10.1097/TA.0b013e3181e97761
Hydroxyapatite-coated pins versus titanium alloy pins in external fixation at the wrist: a controlled cohort study
Abstract
Background: The purpose of this study was to analyze whether the prevalence of pin-related complications can be reduced by the use of hydroxyapatite (HA)-coated pins in external fixators applied for unstable wrist fractures.
Methods: Forty patients (160 pins) were randomized for standard uniplanar fixator treatment with the use of identically designed pins either composed of titanium-alloy (Ti6Al4V) (n = 20) or coated by HA (n = 20). Each pin site was clinically evaluated with regard to erythema, drainage, pain value, and radiologically assessed concerning loosening at T1 (mean, 9 days), T2 (mean, 43 days), and T3 (mean, 56 days). In case of pin-track complication, the patient was followed continuously. The need for antibiotics or additional surgery was documented. Bone mineral density was analyzed by Dual Energy X-ray Absorptiometry. At fixator removal (T2), the pin-extraction strength was measured by the use of a digital-torque-wrench.
Results: Two minor pin-track infections requiring oral antibiotics occurred in the HA-pin group (2.7%) (p > 0.05). The vast majority of clinical pin-site parameters were comparable in both groups. At the end of the fixator therapy, there were 16 loose pins (n(Ti6AL4V-group) = 10; n(HA-group) = 6). The rate of loose pins was correlated to patient's age (p < 0.05) but not to bone mineral density values or the occurrence of pin-site infection. Finally, no significant difference between the two groups was detected with regard to the prevalence of clinical relevant pin-site complications (p = 0.80).
Conclusions: In external fixation of the wrist, the use of HA-coated pins yields no clinical advantages: there is a trend toward a superior pin-bone anchorage, but a tendency of increased susceptibility for minor pin-track infections.
Similar articles
-
Clinical benefit of hydroxyapatite-coated pins compared with stainless steel pins in external fixation at the wrist: a randomised prospective study.Injury. 2010 Oct;41(10):1031-1036. doi: 10.1016/j.injury.2010.03.030. Injury. 2010. PMID: 20444448 Clinical Trial.
-
Titanium alloy pins versus stainless steel pins in external fixation at the wrist: a randomized prospective study.J Trauma. 2008 May;64(5):1275-80. doi: 10.1097/TA.0b013e31815e40e0. J Trauma. 2008. PMID: 18469650 Clinical Trial.
-
Improvement of the bone-pin interface strength in osteoporotic bone with use of hydroxyapatite-coated tapered external-fixation pins. A prospective, randomized clinical study of wrist fractures.J Bone Joint Surg Am. 2001 May;83(5):717-21. doi: 10.2106/00004623-200105000-00010. J Bone Joint Surg Am. 2001. PMID: 11379741 Clinical Trial.
-
Hydroxyapatite-coated external fixation pins.Expert Rev Med Devices. 2005 Jul;2(4):465-71. doi: 10.1586/17434440.2.4.465. Expert Rev Med Devices. 2005. PMID: 16293085 Review.
-
Hydroxyapatite pins for external fixation: is there sufficient evidence to prove that coated pins are less likely to be replaced prematurely?Acta Orthop Traumatol Turc. 2015;49(4):410-5. doi: 10.3944/AOTT.2015.14.0206. Acta Orthop Traumatol Turc. 2015. PMID: 26312469 Review.
Cited by
-
Potential of Titanium Pins Coated with Fibroblast Growth Factor-2-Calcium Phosphate Composite Layers to Reduce the Risk of Impaired Bone-Pin Interface Strength in the External Fixation of Distal Radius Fractures.J Clin Med. 2024 May 22;13(11):3040. doi: 10.3390/jcm13113040. J Clin Med. 2024. PMID: 38892751 Free PMC article.
-
Titanium Kirschner Wires Resist Biofilms Better Than Stainless Steel and Hydroxyapatite-coated Wires: An In Vitro Study.Strategies Trauma Limb Reconstr. 2019 May-Aug;14(2):57-64. doi: 10.5005/jp-journals-10080-1426. Strategies Trauma Limb Reconstr. 2019. PMID: 32742415 Free PMC article.
-
Interventions for treating wrist fractures in children.Cochrane Database Syst Rev. 2018 Dec 19;12(12):CD012470. doi: 10.1002/14651858.CD012470.pub2. Cochrane Database Syst Rev. 2018. PMID: 30566764 Free PMC article.
-
Hydroxyapatite-coated compared with stainless steel external fixation pins did not show impact in the rate of pin track infection: a multicenter prospective study.Int Orthop. 2023 May;47(5):1163-1169. doi: 10.1007/s00264-023-05717-w. Epub 2023 Feb 11. Int Orthop. 2023. PMID: 36773051 Free PMC article. Clinical Trial.
-
Role of coatings and materials of external fixation pins on the rates of pin tract infection: A systematic review and meta-analysis.World J Orthop. 2021 Nov 18;12(11):920-930. doi: 10.5312/wjo.v12.i11.920. eCollection 2021 Nov 18. World J Orthop. 2021. PMID: 34888152 Free PMC article.
Publication types