Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2010 Dec;17(4):285-90.
doi: 10.1177/1553350610379427. Epub 2010 Sep 3.

Evaluation of acute fixation strength for mechanical tacking devices and fibrin sealant versus polypropylene suture for laparoscopic ventral hernia repair

Affiliations

Evaluation of acute fixation strength for mechanical tacking devices and fibrin sealant versus polypropylene suture for laparoscopic ventral hernia repair

Lora Melman et al. Surg Innov. 2010 Dec.

Abstract

Background: The purpose of this comparative study is to evaluate the acute fixation strength of mechanical tacking devices and fibrin sealant against polypropylene suture for laparoscopic ventral hernia repair.

Methods: Three metallic mechanical tacking devices (ProTack, Salute, EndoANCHOR), 4 absorbable tacking devices (AbsorbaTack, PermaSorb, I-Clip, and SorbaFix), and 2 types of fibrin sealant (Tisseel, Artiss) were compared with 0-polypropylene suture. Three constructs from each device or an amount of sealant sufficient to cover a 3 × 3 cm(2) area were used to affix a 4 × 3 cm piece of absorbable barrier-coated mesh (Proceed, Ethicon, Inc) to the peritoneal surface of porcine abdominal wall. Ten samples were completed for each fixation modality. Acute fixation strength was measured via a lap shear test on an Instron tensiometer.

Results: Acute fixation strength was significantly greater for suture (59.7 7.2 N) compared with all laparoscopic tacking devices and to fibrin sealant (P < .001 for all comparisons). Protack (29.5 ± 2.8 N) was stronger than Absorbatack (13.2 ± 3.7 N; P = .029). Protack, Permasorb, SorbaFix, and I-clip were stronger than fibrin sealant (P < .05 for all comparisons).

Conclusions: The acute fixation strengths of metallic or absorbable tacks as well as fibrin sealant are all significantly less than that achieved with polypropylene suture. These factors should be considered in selecting the type of mechanical fixation for patients undergoing laparoscopic ventral hernia repair.